摘 要:详细介绍了Medrad■Intego PET药品自动给药系统的基本组成和工作原理,分析了该系统在使用过程中出现的3例故障的原因并给出具体的故障排除方法,指出了该系统在日常维护保养中应注意的问题,为同行及维修人员提供了重要参考。The main components and working principles of Medrad■Intego PET Infusion System were described in detail.The causes of three cases of faults during the use of the system were analyzed and the countermeasures were proposed accor-dingly.The key points for the routine maintenance of the system were expounded,providing important references for colleagues and maintenance personnel.