摘  要:首先对博科BKQ-B型Ⅱ系列压力蒸汽灭菌器加热保护装置、防干烧报警装置、门安全联锁装置和超温超压保护装置等工作原理讲解,再对故障代码Err064、错误代码Err 032和无规律跳闸等三个故障案例按照故障现象、故障分析、故障排除过程和小结四个环节进行详述,以期广大医院工程师对于该机型有更详尽的了解,便于在实际工作中借鉴使用。This paper on the brocade BKQ-B type Ⅱ series pressure steam sterilizer heating protection device,prevent dry alarm device,safety interlock and super pressure protection device etc.The working principle,fault codes such as Err064,Err032 and irregular trip three failure case according to the failure phenomenon,failure analysis,troubleshooting,and detailed summary four link,to make the hospital engineer for the aircraft have a more detailed understanding,easy to use for reference in the practical work.