基于改造氨氮模块的连续流动分析仪检测水中六价铬含量  被引量:3

Determination of Hexavalent Chromium Content in Water by Continuous Flow Analyzer Based on Modified Ammonia Nitrogen Module

在线阅读下载全文

作  者:任淑艳[1] 栾田[1] 马可佳 赵岚[1] REN Shuyan;LUAN Tian;MA Kejia;ZHAO Lan(Heilongjiang Center for Disease Control and Prevention,Harbin 150030,China)

机构地区:[1]黑龙江省疾病预防控制中心,黑龙江哈尔滨150030

出  处:《现代食品》2022年第4期222-224,共3页Modern Food

摘  要:目的:将连续流动分析仪氨氮模块进行简单改造,用于检测水中六价铬含量。方法:采用SKALAR San++型连续流动分析仪,根据六价铬在酸性条件下与二苯碳酰二肼反应生成一种紫红色络合物,在540 nm处检测六价铬含量。结果:六价铬在0~0.2 mg·L^(-1)具有良好的线性关系,相关系数为0.9993;检出限为0.001 mg·L^(-1);相对标准偏差<3%,回收率在95.1%~103.0%;本法和国标二苯碳酰二肼分光光度法比较无显著性差异。结论:改装后的连续流动分析仪测定水中六价铬具有检测速度快、准确度和灵敏度高、对环境污染小的特点,适用于大批量水样的分析,同时减少了实验室二次采购仪器的成本。Objective:The ammonia nitrogen module of continuous flow analyzer was simply modified to detect hexavalent chromium content in water.Method:SKALAR SAN++continuous flow analyzer was selected.According to the reaction of hexavalent chromium with diphenylcarbazide under acidic conditions to form a purplish red complex,and the content of hexavalent chromium was determined at 540 nm.Result:This methed has a good relativity between 0 mg·L^(-1)and 0.2 mg·L^(-1).The correlation coefficient was 0.9993 and the detection limit was 0.001 mg·L^(-1).The relative standard deviation was<3%,and the recovery rates was 95.1%~103.0%.It was not significant difference compared with standard method used with diphenylcarbazide for spectrophotometry.Conclusion:The modified continuous flow analyzer method for the determination of hexavalent chromium in water has the characteristics of fast detection speed,high accuracy and sensitivity,and little environmental pollution.It can be applied to analyse enormous water samples and which has saved the costs of making the second purchase of equipment for laboratories.

关 键 词:连续流动分析仪 氨氮通道 六价铬 仪器改造 

分 类 号:R123.1[医药卫生—环境卫生学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象