作 者:霍文婷 邵佳欢 HUO Wen-ting;SHAO Jia-huan(Tianjin Binhai New Area inspection and Testing Center,Tianjin 300450,China)
机构地区:[1]天津市滨海新区检验检测中心,天津300450
出 处:《品牌与标准化》2022年第2期38-40,共3页
摘 要:紫外可见分光光度计在使用过程中,由于多种原因会引起波长或杂散光检定误差,导致测量结果不准确,为了得到更为准确的实验结果,必须进一步控制紫外可见分光光度计检定误差。本文结合作者工作经验,以紫外可见分光光度计检定原理作为切入点,对常见误差来源进行分析,并对如何有效控制此类误差进行了讨论。During the use of UV-Vis spectrophotometer,the verification error of wavelength or stray light will be caused due to various reasons,resulting in inaccurate measurement results. In order to obtain more accurate experimental results,the verification error of UVVis spectrophotometer must be further controlled. Combined with my own working experience,taking the verification principle of UVVis spectrophotometer as the starting point,this paper analyzes the common error sources,and discusses how to effectively control this kind of error.