检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:袁子龙 刘涛[2] 张彪 陈迢 李翠玲 张照喜[2] 吴磊 Yuan Zilong;Liu Tao;Zhang Biao;Chen Tiao;Li Cuiling;Zhang Zhaoxi;Wu Lei(Department of Clinical Medicine,Medical College,Wuhan University of Science and Technology,Wuhan,430065,China;Department of Radiology,Hubei Cancer Hospital,Tongji Medical College,Huazhong University of Science and Technology,Wuhan 430079,China;Department of Radiological Health,Wuhan Hospital for the Prevention and Treatment of Occupational Disease,Wuhan 430015,China)
机构地区:[1]武汉科技大学医学院临床医学系,武汉430065 [2]湖北省肿瘤医院放射科,武汉430079 [3]武汉市职业病防治院放射卫生科,武汉430015
出 处:《中华放射医学与防护杂志》2022年第3期225-229,共5页Chinese Journal of Radiological Medicine and Protection
摘 要:目的探讨能谱纯化技术(SPS)、器官剂量调制技术(OBTCM)以及两者结合对婴幼儿头部CT扫描器官剂量及图像质量的影响。方法利用两个头部仿真体模(CIRS 1岁和5岁),分别采用参考扫描模式(Reference)、Reference+OBTCM、SPS及SPS+OBTCM 4种模式进行扫描;测量并比较不同体模不同扫描模式眼晶状体、脑部前侧及脑部后侧的辐射剂量,眼眶和脑实质区域的噪声水平及对比噪声比(CNR)。结果相比Reference模式,1岁与5岁体模的眼晶状体剂量在Reference+OBTCM模式分别减少约(21.89±0.01)%和(28.33±0.34)%;在SPS模式下,1岁和5岁体模的眼晶状体剂量减少分别为(71.38±1.30)%和(53.72±2.42)%;SPS+OBTCM模式分别减少约(71.12±2.54)%和(55.73±1.90)%。不同体模不同扫描模式眼眶及脑实质的噪声水平,差异有统计学意义(F=5.67~85.47,P<0.05);与Reference模式相比,Reference+OBTCM模式的噪声水平在不同体模不同部位中略有增加(<1.45HU),而SPS和SPS+OBTCM模式产生较小的噪声增加(<2.58 HU);不同体模不同扫描模式不同部位的CNR差异无统计学意义(P>0.05)。结论SPS和SPS+OBTCM模式可以显著降低婴幼儿头部CT扫描中眼晶状体及整个图像平面的辐射剂量,同时保持图像质量。Objective To compare the dose and image quality of selected photon shield(SPS)technique,organ-based tube current modulation(OBTCM)technique and the combination of these two techniques for reducing the organ dose in head CT examination for infants.Methods Two anthropomorphic head phantoms(CIRS 1-yr-old and 5-yr-old)were scanned by using Reference mode,Reference+OBTCM mode,SPS mode and SPS+OBTCM mode,respectively.Radiation doses to the lens of the eye,the anterior of the brain,the posterior of the brain,noise level and CNR of orbit and brain in different phantoms were measured and compared by using different scanning modes.Results Compared with Reference mode,the doses to the lens of the eye in 1-yr-old and 5-yr-old phantom decreased by(21.89±0.01)%and(28.33±0.34)%,respectively.In SPS mode,the reduction in doses to the lens of the eye in 1-yr-old and 5-yr-old phantom were(71.38±1.30)%and(53.72±2.42)%,respectively.In SPS+OBTCM mode,the reduction was(71.12±2.54)%and(55.73±1.90)%,respectively.There was significant difference in the noise level of orbit and brain in different phantoms under various scanning modes(F=5.67-85.47,P<0.05).The noise level in OBTCM mode compared with reference mode increased slightly(<1.45 HU)in various phantoms.SPS and SPS+OBTCM mode resulted in a small noise increase(<2.58 HU).There was no significant difference in CNR of different phantoms under various scanning modes(P>0.05).Conclusions SPS and SPS+OBTCM mode can significantly reduce the radiation dose of lens and the whole image plane in the head CT scan for infants,with maintaining the image quality.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.200