飞利浦DR胸片架移动类故障维修3例  被引量:3

Maintenance of motion faults of wall stand of PHILIPS DR:three case reports

在线阅读下载全文

作  者:冉华锋 刘传高[1] RAN Hua-feng;LIU Chuan-gao(Department of Equipment,the 2nd Affiliated Hospital of Bengbu Medical College,Bengbu 233040,Anhui Province,China)

机构地区:[1]蚌埠医学院第二附属医院医疗设备科,安徽蚌埠233040

出  处:《医疗卫生装备》2022年第4期102-104,108,共4页Chinese Medical Equipment Journal

摘  要:介绍了飞利浦DR胸片架系统的结构组成,详细分析了胸片架不能移动、胸片架移动缓慢、球管突发自主移动3例故障产生的原因,并提出了具体的排除方法,为从事医疗设备维修的工程师排除此类故障提供了参考。The composition of the wall stand of PHILIPS DR was introduced.A detailed analysis of the causes of three cases of motion faults,namely,the wall stand could not be moved,the stand moved slowly and the sudden autonomous movement of the ball tube,was conducted,and some countermeasures were proposed to provide references for medical equipment maintenance engineers to treat similar failures.

关 键 词:DR 胸片架 移动类故障 故障维修 

分 类 号:R197.39[医药卫生—卫生事业管理]

 