检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:赵东升[1] 孔康辉 ZHAO Dongsheng;KONG Kanghui(Department of Medical Equipment,Henan Provincial People’s Hospital,Zhengzhou Henan 450003,China)
机构地区:[1]河南省人民医院医学装备部,河南郑州450003
出 处:《中国医疗设备》2022年第5期178-181,共4页China Medical Devices
摘 要:本文整理了两例UHI-3腹腔镜气腹机电源故障和主控板故障的维修情况,通过电路板电路图还原,进行电路分析,分别通过电源板更换7815芯片、主控板更换模拟开关DG408DY芯片的操作,成功实现了芯片级维修,有效地节省了维修费用,降低运行成本。This paper analyzed two cases of UHI-3 laparoscopic pneumoperitoneum machine and electrical source fault and maintenance situation of main control board fault.Through reduction by PCB circuit diagram restoration,the circuit analysis,respectively by the replacement of 7815 chips power panel,main control board to replace analog switch DG408DY chips,the chip level maintenance was implemented.It saves the maintenance cost effectively,reduces the operation cost.
关 键 词:UHI-3腹腔镜气腹机 芯片级维修 7815芯片 DG408DY芯片 还原电路板电路图
分 类 号:R197.39[医药卫生—卫生事业管理] TH776.1[医药卫生—公共卫生与预防医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.30