扫描镜热辐射校正对FY-4A星AGRI定标精度的提升研究  被引量:1

Study on the improvement of scan mirror thermal radiation correction on the calibration accuracy of FY-4A AGRI

在线阅读下载全文

作  者:李秀举 王保勇 吴亚鹏 韩昌佩[1,2] 曹琦 周树添 王伟成 李盼盼[1,2] LI Xiu-Ju;WANG Bao-Yong;WU Ya-Peng;HAN Chang-Pei;CAO Qi;ZHOU Shu-Tian;WANG Wei-Cheng;LI Pan-Pan(Key Laboratory of Infrared Detection and Imaging Technology,Chinese Academy of Sciences,Shanghai 200083,China;Shanghai Institute of Technical Physics,Chinese Academy of Sciences,Shanghai 200083,China)

机构地区:[1]中国科学院红外探测与成像技术重点实验室,上海200083 [2]中国科学院上海技术物理研究所,上海200083

出  处:《红外与毫米波学报》2022年第3期608-617,共10页Journal of Infrared and Millimeter Waves

基  金:中国科学院上海技术物理研究所创新专项基金(CX-208)。

摘  要:多通道扫描成像辐射计(Advanced Geostationary Radiation Imager,AGRI)是风云四号A星的主要有效载荷之一。为了满足AGRI长波红外遥感数据高精度定量化应用的要求,构建了扫描镜热辐射随机械转角变化的函数模型,提出了一种可用于星上黑体和地球场景观测遥感数据的校正算法。基于FY-4A星AGRI的在轨数据,分析了模型参数对扫描镜温度的依赖性,研究评估了校正算法对长波红外定标精度的改善效果。在扫描镜温度场平稳时间段内,扫描镜热辐射校正后,B11(8.0~9.0μm)、B12(10.3~11.3μm)、B13(11.5~12.5μm)波段的定标偏差分别可以得到-2.81 K~+1.06 K、-0.60 K~+0.19 K、-0.68 K~+0.24 K的显著改善。与IASI交叉定标检验结果表明,改善后的在轨定标亮度温度偏差优于0.5 K@290 K。The Advanced Geostationary Radiation Imager(AGRI)is one of the main payloads of Fengyun-4A(FY-4A).In order to satisfy requirements of high-accuracy quantitative application of AGRI long-wave infrared(LWIR)remote sensing data,a function model of scan mirror thermal radiation changing with mechanical rotation angle is constructed,and a correction algorithm for remote sensing data of on-board blackbody and earth scene observation is presented.Based on the on-orbit data of FY-4A AGRI,the dependence of the model parameters on the scan mirror temperature is analyzed,and the improvement effect of the correction algorithm on the calibration accuracy of LWIR is studied and evaluated.After thermal radiation correction of scan mirror in the stationary period of mirror temperature field,the calibration deviation of B11(8.0~9.0μm),B12(10.3~11.3μm),B13(11.5~12.5μm)can be improved significantly by-2.81 K~+1.06 K,-0.60 K~+0.19 K,-0.68 K~+0.24 K respectively.The result of inter-calibration validation with IASI shows that the improved on-orbit calibration brightness temperature bias is better than 0.5 K@290 K.

关 键 词:热辐射校正 定标 仪器背景辐射 扫描镜 AGRI 风云四号 

分 类 号:TP732.2[自动化与计算机技术—检测技术与自动化装置]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象