检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:毕欣雨 于健伟 许玥 韩莹瑶 羊月祺[1] Bi Xinyu;Yu Jianwei;Xu Yue;Han Yingyao;Yang Yueqi(Department of Clinical Medical Engineering,Jiangsu Province Hospital,Nanjing 210029,China)
机构地区:[1]江苏省人民医院临床医学工程处,南京210029
出 处:《现代仪器与医疗》2022年第3期87-89,93,共4页Modern Instruments & Medical Treatment
摘 要:目的回顾并总结新生儿黄疸治疗箱常见故障的处置方法并制定针对性预防性维护方案.方法统计某医院2016-2019年新生儿黄疸治疗箱的故障情况,介绍频发故障的解决方案.结合故障处置和使用情况,制定预防性维护方案并执行.结果新生儿黄疸治疗箱的频发故障主要有光路故障,风道故障和温度探头故障三大类.针对这三类故障制定预防性维护方案并执行后,故障数量明显下降,医疗安全得到了有效保障.结论通过总结新生儿黄疸治疗箱常见故障的排查和维修方法,针对性开展预防性维护,可以维持设备稳定运行,保证治疗有效性,提高医疗效率,确保患儿人身安全.Objective To review and summarize the treatment methods of frequent failures of neonatal jaundice treatment boxes and allow the development of more tailored programs.Methods From 2016 to 2019,the failures of neonatal jaundice treatment boxes in a hospital were counted,the solutions to frequent failures were analyzed.The preventive maintenance plans were formulated and implemented in combination with failure handling and usage.Results The frequent failures of neonatal jaundice treatment box mainly include optical path failure,air duct failure and temperature probe failure.After the preventive maintenance plan was formulated and implemented for these three types of faults,the number of faults decreased significantly,and medical safety was effectively guaranteed.Conclusion By summarizing the troubleshooting and maintenance methods of its common failures,which carrying out preventive maintenance,the stable operation of the equipment can be maintained and the effectiveness of the treatment can be guaranteed.It can also improve the efficiency of medical treatment,and ensure the children personal safety.
关 键 词:新生儿黄疸治疗箱 故障分析 预防性维护 MBI1816芯片
分 类 号:R197.32[医药卫生—卫生事业管理] R454.2[医药卫生—公共卫生与预防医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.3