基于UF1000i尿沉渣分析仪对Clinitek Atlas尿干化学分析仪白细胞计数与反射率的关系研究及思考  被引量:4

Research and Thinking on the Relationship between White Blood Cell Count and Reflectance of Clinitek Atlas Urine Dipstick Chemistry Analyzer Based on UF1000i Urine Sediment Analyzer

在线阅读下载全文

作  者:张洪波[1] 张辉 黄铭 陈星 李果 ZHANG Hong-bo;ZHANG Hui;HUANG Ming;CHEN Xing;LI Guo(Department of Laboratory Medicine,Tongji Hospital,Tongji Medical College,Huazhong University of Science and Technology,Wuhan 430030,China)

机构地区:[1]华中科技大学同济医学院附属同济医院检验科,武汉430030

出  处:《现代检验医学杂志》2022年第4期165-168,共4页Journal of Modern Laboratory Medicine

摘  要:目的借助UF1000i尿沉渣分析仪,验证尿干化学分析仪中白细胞计数和反射率之间的潜在关系,从而为尿干化学分析仪的定量改进提供新的思路。方法通过筛选尿白细胞计数大于1000/μl的尿标本,在UF1000i尿沉渣分析仪和Clinitek Atlas尿干化学分析仪上同时测试,记录尿沉渣白细胞计数和尿干化学反射率值,并统计分析两者的相关定性等级的反射率临界点和白细胞临界点验证。结果尿沉渣白细胞计数(X)和反射率(Y)存在较好的关系曲线:Y=109.14 ln(X)+931.94,r=0.9958;尿干化学微量、1+,2+和3+的临界值反射率分别为1150,1231,1380和1553;对应的白细胞临界值(个/μl)分别为5,10,52和330。结论尿干化学白细胞计数与反射率的确立,为现行的尿干化学分析仪的校准提供了新的解决思路,也为尿干化学分析仪由定性向定量发展奠定了基础,从而使尿干化学检测服务于临床的质量更进一步。Objective To verify the potential relationship between the white blood cell(WBC)count and reflectance in urine dipstick chemistry analyzer by means of UF1000i urine sediment analyzer,so as to provide a new idea for quantitative improvement of urine dipstick chemstry analyzer.Methods By screening suitable urine samples(WBC count>1000/μl)and testing them simultaneously on UF1000i urine sediment analyzer and Clinitek Atlas urine chemical analyzer,the WBC count and urine dipstick chemistry reflectance of urine sediment were recorded,and the correlation between them was calculated.Results The curve between the urine WBC count and reflectance was good:Y=109.14 ln(X)+931.94,and the value of r was 0.9958.The critical reflectance of trace,1+,2+and 3+were 1150,1231,1380 and 1553,respectively.The corresponding WBC count(/μl)were 5,10,52 and 330,respectively.Conclusion The establishment of the WBC count and reflectivity of the urine dipstick chemical analyzer provides a new solution for the calibration of current urine dipstick chemistry analyzer,and also lays a foundation for the development of urine dipstick chemistry analyzer from qualitative to quantitative,so as to further improve the clinical quality of urine dipstick chemistry test.

关 键 词:尿干化学分析仪 反射率 中性粒细胞脂肪酶 白细胞脂肪酶 校准验证 

分 类 号:R446.12[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象