TIPS术后支架评估中的光谱CT最佳单能级图像研究  

Optimal single-energy image of spectral CT in stent evaluation after TIPS

在线阅读下载全文

作  者:汤朝晖 张建生[1] 陈其龙 邓薇薇 王劲[1] Tang Zhaohui;Zhang Jiansheng;Chen Qilong;Deng Weiwei;Wang Jin(Department of Radiology,the Third Affiliated Hospital of Sun Yat-sen University,Guangzhou 510630,China;Clinical Research Department,Philips Healthcare,Shanghai 200072,China)

机构地区:[1]中山大学附属第三医院放射科,广州510630 [2]飞利浦医疗临床科研部,上海200072

出  处:《中华肝脏外科手术学电子杂志》2022年第5期463-468,共6页Chinese Journal of Hepatic Surgery(Electronic Edition)

摘  要:目的探讨IQon双层探测器光谱CT在经颈静脉肝内门体分流术(TIPS)后支架评估中的最佳单能级图像。方法回顾性分析2020年3月至2020年9月在中山大学附属第三医院行上腹部光谱CT检查的30例门静脉高压症TIPS术后随访患者影像学资料。其中男18例,女12例;年龄38~72岁,中位年龄46岁。患者均签署知情同意书,符合医学伦理学规定。采用门静脉期光谱CT数据在后处理工作站重建出40~100 keV(间隔10 keV)、100~200 keV(间隔20 keV)单能级图像。在常规图像和12组单能级图像上进行支架曲面重建,测量管腔内及同层竖脊肌的CT值和噪声值。计算13组图像的对比噪声比(CNR)。两名10年以上诊断经验的医师对图像的腔内伪影程度及诊断信心进行主观评分。图像质量评价指标比较采用单因素方差分析和LSD-t检验,或Friedman检验。图像质量主观评价的一致性采用Kappa检验。结果随着能级降低,TIPS支架管腔内CT值逐渐升高,其中80~200 keV能级图像的CT值均明显低于常规图像(LSD-t=-2.6,-5.3,-7.1,-9.2,-10.4,-11.1,-11.6,-12.1;P<0.05);40~60 keV能级图像的管腔内CT值明显高于常规图像(LSD-t=36.7,18.5,7.8;P<0.05)。随着能级降低,管腔内SD值逐渐升高,其中80~200 keV能级图像的SD值均明显低于常规图像(LSD-t=-4.4,-5.1,-6.0,-6.6,-6.1,-6.3,-6.3,-6.3;P<0.05);40 keV能级图像的管腔内SD值明显高于常规图像(LSD-t=4.5,P<0.05)。随着能级降低,支架的CNR逐渐升高。其中90~200 keV能级图像的支架CNR值均明显低于常规图像(LSD-t=-2.2,-3.6,-5.4,-6.1,-6.0,-6.1,-6.2;P<0.05);40~70 keV能级图像的支架CNR值均明显高于常规图像(LSD-t=22.7,16.7,8.4,3.4;P<0.05)。各组图像诊断信心评分显示,90~140 keV组的主观评分均明显高于常规图像组(χ^(2)=292.1,302.8,343.7,270.8;P<0.05),其中120 keV组评分最高,两位医师的主观评分一致性好(κ=0.81)。腔内伪影主观评分显示,90~200 keV组的主观评分均明显高于常规�Objective To investigate the optimal single-energy image of IQon dual-layer spectral detector CT in stent evaluation after transjugular intrahepatic portosystemic shunt(TIPS).Methods Imaging data of 30 patients with portal hypertension who underwent spectral CT scan of the upper abdomen in the Third Affiliated Hospital of Sun Yat-sen University from March to September 2020 were retrospectively analyzed.Among them,18 patients were male and 12 female,aged from 38 to 72 years,with a median age of 46 years.The informed consents of all patients were obtained and the local ethical committee approval was received.The single-energy images of 40-100 keV(interval of 10 keV)and 100-200 keV(interval of 20 keV)were reconstructed at the post-processing workstation using the spectral CT data of portal venous phase.The stent curved planar reconstruction was performed with the conventional images and 12 groups of single-energy images.The CT value and noise value in the lumen and erector spinae of the same layer were measured.The contrast-to-noise ratio(CNR)of images of 13 groups was calculated.The degree of intraluminal artifacts and diagnostic confidence were scored subjectively by 2 radiologists with more than 10 years'experience.The image quality evaluation indexes were compared by one-way ANOVA,LSD-t test or Friedman test.The consistency of subjective evaluation of image quality was assessed by Kappa test.Results With the decrease of energy level,the intraluminal CT value of TIPS stent was gradually increased.The CT values of 80-200 keV images were significantly lower than those of conventional images(LSD-t=-2.6,-5.3,-7.1,-9.2,-10.4,-11.1,-11.6,-12.1;P<0.05).The intraluminal CT values of 40-60 keV images were significantly higher than those of conventional images(LSD-t=36.7,18.5,7.8;P<0.05).With the decrease of energy level,the intraluminal SD value was gradually increased.The SD values of 80-200 keV images were significantly lower than those of conventional images(LSD-t=-4.4,-5.1,-6.0,-6.6,-6.1,-6.3,-6.3,-6.3;P<0.05).The int

关 键 词:计算机体层摄影 光谱法 单能级图像 门体分流术 经颈静脉肝内 支架 高血压 门静脉 

分 类 号:R657.3[医药卫生—外科学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象