钇-90韧致辐射SPECT平面显像能窗和准直器的选择  被引量:2

Selection of energy window and collimator for SPECT bremsstrahlung planar imaging with yttrium-90

在线阅读下载全文

作  者:薛建军[1] 王岐 许惠 谢欣 张月敏 张宇辰[1] 王子馨 汪源 杨爱民[1] XUE Jianjun;WANG Qi;XU Hui;XIE Xin;ZHANG Yuemin;ZHANG Yuchen;WANG Zixin;WANG Yuan;YANG Aimin(Department of Nuclear Medicine,the First Afiliated Hospital of Xi'an Jiaotong University,Xi'an 710061,China;Shaanai Nuclear and Radiation Safety Supervision Station,Xi'an 710054,China;Shaanci Provincial Department of Ecology and Environment,Xi'an710006,China)

机构地区:[1]西安交通大学第一附属医院核医学科,陕西西安710061 [2]陕西省核与辐射安全监督站,陕西西安710054 [3]陕西省生态环境厅,陕西西安710006

出  处:《实用放射学杂志》2022年第9期1533-1537,共5页Journal of Practical Radiology

摘  要:目的研究准直器及能窗对钇-90(^(90)Y)韧致辐射SPECT平面显像图像质量的影响.方法获取90Y韧致辐射能峰,比较低能高分辨率准直器(LEHRC)、高能通用型准直器(HEGPC)及不同能窗时^(90)Y韧致辐射平面显像的灵敏度、源计数/有效视野计数百分比(S/FOV%)、靶/本比值(TBR).结果90Y韧致辐射能谱为连续曲线,在无准直器、LEHRC及HEGPC时能峰分别为83.2 keV、75.3 keV及72.7 keV.HEGPC图像清楚;90 keV±20%、110 keV±20%、80 keV±30%、100 keV±30%及94 keV+40%能窗时,HEGPC灵敏度明显优于LEHRC.任一能窗,HEGPC的TBR及S/FOV%均明显优于LEHRC;10%窗宽,60 keV、70 keV、100 keV、110 keV时,TBR较高,但灵敏度较低;HEGPC时70 keV±20%、100 keV±30%、94 keV±40%窗宽,TBR分别为59.53、30.91、40.53,灵敏度分别为18.11%、55.67%、59.09%.结论HEGPC的90Y韧致辐射图像质量优于LEHRC;94 keV+40%是90Y韧致辐射显像的最佳能窗.Objective To study the effect of energy window and collimator on the image quality of yttrium-9o(o Y)bremsstrahlung SPECT planar imaging.Methods The energy peak of 90 Y bremsstrahlung were obtained,and the sensitivity,percentage of the source counts in useful field of view(S/FOV%)and target-background-ratio(TBR)of 90 Y bremsstrahlung planar imaging at low energy high resolution collimator(LEHRC)and high energy general purpose collimator(HEGPC)were compared,respectively.Results The energy spectrum of 90 Y bremsstrahlung was a continuous curve with peaks of 83.2 keV,75.3 keV and 72.7 keV without collimator,with LEHRC and HEGPC,respectively.The images obtained with the HEGPC were clear.The sensitivity of HEGPC was significantly better than that of LEHRC at 90 keV±20%,110 keV±20%,80 keV±30%,100 keV±30%and 94 keV±40%energy windows.The TBR and S/FOV%of HEGPC were significantly better than those of LEHRC at any energy window.At 10%window width,the TBR was higher at 60keV,70keV,100keV,110keV,but the sensitivity was lower;at 70keV±20%,100keV±30%,94 keV±40%energy windows with HEGPC,TBR was 59.53,30.91,40.53,and sensitivity was 18.11%,55.67%,59.09%,respectively.Conclusion The 90Y bremsstrahlung image quality of HEGPC is better than that of LEHRC;94 keV±40%is the best energywindow.

关 键 词:钇-90 韧致辐射 准直器 能窗 

分 类 号:R817-3[医药卫生—影像医学与核医学] R814.42[医药卫生—放射医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象