AU5831自动生化分析仪2种试剂检测唾液酸化糖链抗原-6临床应用评价  被引量:1

Clinical evaluation of AU5831 automatic biochemical analyzer for Krebs Von den Lungen-6 kit

在线阅读下载全文

作  者:彭定辉[1] 任玉国 庄伟 廖凡路 唐鹏[1] 陈薇[1] PENG Dinghui;REN Yuguo;ZHUANG Wei;LIAO Fanlu;TANG Peng;CHEN Wei(Laboratory Department of Central South Hospital of Wuhan University,Wuhan,430071,China;Jinan People's Hospital)

机构地区:[1]武汉大学中南医院检验科,武汉430071 [2]济南市人民医院

出  处:《临床血液学杂志》2023年第2期121-124,共4页Journal of Clinical Hematology

摘  要:目的:运用AU5831自动生化分析仪和2种试剂检测唾液酸化糖链抗原-6(KL-6),评价试验产品与参比产品的临床使用效能。方法:选取300例研究标本,其中间质性肺疾病患者标本75例,无间质性肺疾病病症患者标本205例,干扰病例样本20例,使用2种试剂的胶乳免疫比浊法和胶乳凝集法对研究样本进行平行盲法检测,比较试验产品与参比产品的检测结果,进行离群值检验、回归分析、Bland-Altman一致性分析及医学决定水平的预期偏倚分析。结果:试验产品与参比产品各检测KL-6 300例。根据NCCLS EP9-A2文件进行方法间离群值检查,以绝对差值和相对差值的均值4倍为限值,数据均未超出检测限(TLE)数值,无离群值。Deming回归方程为Y=0.907 9+0.995 1X(Pearson相关系数:0.999 9,95%CI 0.999 9~0.999 9),满足评价标准。Bland-Altman分析,一致性界限外无散点。试验产品与参比产品的一致性检验发现κ=0.965,认为试验产品与参比产品一致性良好。结论:AU5831自动生化分析仪检测KL-6的2种试剂盒在临床应用中具有可比性,一致性较好,可用于临床检测。Objective: AU5831 automatic biochemical analyzer and two reagents were used to detect Krebs Von den Lungen-6(KL-6), and to evaluate the clinical efficacy of the test kit and reference product. Methods: A total of 300 research specimens were selected in our hospital, including 75 specimens from patients with interstitial lung disease, 205 specimens from patients without interstitial lung disease symptoms, and 20 specimens from interference cases. Latex immunoturbidimetry and latex agglutination were used to detect the clinical samples by parallel blind method. Based on the experiment results, outlier test, regression analysis, Bland-Altman consistency analysis, expected bias analysis at the level of medical decision and consistency analysis of qualitative results were selected for the study. Results: KL-6 was detected in 300 samples of test product and reference product respectively. According to NCCLS EP9-A2 document, the outliers between methods were checked which was limited to 4 times the mean value of absolute difference and relative difference, and no outlier was found. Deming regression equation was develop as: Y=0.907 9+0.995 1X with pearson correlation coefficient: 0.999 9. The 95% confidence interval was 0.999 9-0.999 9. There were no scattered points outside the limits of agreement. The consistency test of the test product and the reference product found that κ=0.965, and the consistency between the test product and the reference product was considered to be good. Conclusion: The two kits for the detection of KL-6 by the AU5831 automatic biochemical analyzer are comparable in clinical application and have good consistency, which can be used for clinical detection.

关 键 词:唾液酸化糖链抗原-6 胶乳免疫比浊法 间质性肺炎 临床评价 

分 类 号:R457.1[医药卫生—治疗学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象