作 者:胡崇庆 袁杰[1] 白雪[1] 孙训[1] Hu Chongqing;Yuan Jie;Bai Xue;Sun Xun(Shanghai Institute of Measurement and Testing Technology)
机构地区:[1]上海市计量测试技术研究院
出 处:《上海计量测试》2022年第6期23-25,28,共4页Shanghai Measurement and Testing
摘 要:对目前常用的18种规格型号的个人剂量当量监测仪做剂量当量率和剂量当量测量性能实验,结果显示:不论采用何种测量方式,在其中一种测量模式下相对固有误差满足规程技术要求,另一种测量模式下相对固有误差也能满足规程的技术要求,由此除指定用于剂量当量测量外,个人剂量当量H_(P)(10)监测仪一般只需按照剂量率测量模式进行检定。Through the experiment on the dose equivalent rate and dose equivalent measurement performance of 18 types of personal dose equivalent monitors currently used,the results show that the relative intrinsic error in one mode meets the technical requirements of the procedure,and the relative intrinsic error in the other measurement mode also meets the technical requirements of the procedure,regardless of the measurement method used,the personal dose equivalent H_(P)(10)monitor generally only needs to be calibrated according to the dose rate measurement mode.
关 键 词:相对固有误差 表面下10 mm处的剂量当量监测仪 个人剂量当量 剂量当量率 剂量当量(率)响应
分 类 号:R144[医药卫生—公共卫生与预防医学]