基于二极管泵浦固体激光器的便携式LIBS系统设计  

Quantitative analysis of element V in iron using portable LIBS based on diode pumped solid-state laser

在线阅读下载全文

作  者:马洪华 李孜涛 袁梦甜 李墨婷 邓凡 陈光辉 李文鑫 余华清 曾庆栋 MA Honghua;LI Zitao;YUAN Mengtian;LI Moting;DENG Fan;CHEN Guanghui;LI Wenxin;YU Huaqing;ZENG Qingdong(School of Physics and Electronic-information Engineering,Xiaogan Hubei 432000,China;Wuhan National Laboratory for Optoelectronics(WNLO),Huazhong University of Science and Technology,Wuhan 430074,China;Faculty of Physics and Electronic Science,Hubei University,Wuhan 430062,China)

机构地区:[1]湖北工程学院物理与电子信息工程学院,湖北孝感432000 [2]华中科技大学武汉光电国家研究中心,武汉430074 [3]湖北大学物理与电子科学学院,武汉430062

出  处:《激光杂志》2023年第3期24-27,共4页Laser Journal

基  金:国家自然科学基金(No.61705064);湖北省自然科学基金(No.2021CFB607);湖北省教育厅优秀中青年科技创新团队计划项目(No.T201617)资助。

摘  要:为实现野外或工业现场等环境对钢材中微量元素V的便携化高精度检测,本文采用二极管泵浦固体激光器作为光源,设计搭建了便携式激光诱导击穿光谱系统。基于攀钢生铁标准样品(GSB03-2582-2010系列)中钒(V)437.92 nm谱线强度和信背比对系统进行参数优化,得出最佳脉冲能量为40 mJ,最佳延时时间为4μs。在最优条件下,采用定点打靶对样品中V元素进行定量分析,定标曲线拟合度决定系数分别为0.995,V元素绝对误差的平均值为0.014%,平均相对误差为6.062%。结果表明,采用小型高脉冲能量二极管泵浦固体激光器配合小型CCD阵列光谱仪搭建的便携式LIBS系统对元素进行快速检测,达到传统LIBS系统的精度水平。该研究为野外或工业现场环境作业场合下的元素快速检测提供了一种新的思路。A portable laser-induced breakdown spectroscopy system based on diode pumped solid-state laser was developed to accomplish the rapid analysis of trace element in steel,and provide high precision and portable rapid detection for field operations or industrial field scenarios.The element V 437.92 nm in samples of iron was selected to optimize the parameters of the portable system.The results show that the optimal pulse energy is 40 mJ and the optimal delay time is 4μs.Under these conditions,V in the sample was quantitatively analyzed by fixed-point shooting and moving shooting respectively.It was obtained that the determination coefficient R2 of calibration curve was 0.995 and 0.991,and the mean of the absolute error of the V element was 0.014%and 0.02%,respectively.The average relative errors were 6.062%and 17.973%,respectively.The results show that the portable LIBS system,which is built by a micro diode pumped laser with high pulse energy and a micro-CCD array spectrometer,can achieve the precision level of the traditional LIBS system for the rapid detection of elements.Meanwhile,the fixed-point beating method has better precision and accuracy than the moving beating method.The research results provide a new idea for the rapid detection of elements in the field environment or in the frequency-changing mobile operations.

关 键 词:二极管泵浦激光器 激光诱导击穿光谱 参数优化 定量分析 快速检测 

分 类 号:O433[机械工程—光学工程]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象