The application study of dual-energy CT nonlinearblending technique in pulmonary angiography  

在线阅读下载全文

作  者:Siqi Yi Peng Zhou Yakun He Changjiu He Shibei Hu 

机构地区:[1]Department of Radiology,Sichuan Cancer Hospital&Institute,Sichuan Cancer Center,University of Electronic Science and Technology of China,Chengdu 610041,China

出  处:《Oncology and Translational Medicine》2023年第1期22-27,共6页肿瘤学与转化医学(英文版)

基  金:Supported by a grant from the Science and Technology Plan of Sichuan Province(No.2021YFS0225);the Science and Technology Plan of Chengdu(No.2021-YF05-01507-SN).

摘  要:Objective This study aimed to explore the feasibility of enhancing image quality in computed tomography(CT) pulmonary angiography (CTPA) and reducing radiation dose using the nonlinear blending (NLB)technique of dual-energy CT.Methods A total of 61 patients scheduled for CTPA were enrolled, and 30 patients underwent dual-energyscanning. Nonlinear blending images (NLB group) and three groups of linear blending images (LB group,80 kV group, and 140 kV group) were reconstructed after scanning;31 patients underwent single-energyscanning (120 kV group). The CT values and standard deviations of the pulmonary trunk, left and rightpulmonary arteries, and ipsilateral back muscle at the bifurcation level of the left and right pulmonaryarteries were measured. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the fivegroups were calculated. The subjective image quality of the five groups was assessed. The radiation dosesof the dual- and single-energy groups were recorded and calculated.Results The CNR and SNR values of blood vessels in the NLB group were significantly higher than thosein the LB, 140 kV, and 80 kV groups (CNR of pulmonary artery trunk: t = 3.50, 4.06, 7.17, all P < 0.05;SNRof pulmonary trunk: t = 3.76, 4.71, 6.92, all P < 0.05). There were no statistical differences in the CNR andSNR values between the NLB group and 120 kV group (P > 0.05). The effective radiation dose of the dualenergygroup was lower than that of the single-energy group (t = –4.52, P < 0.05). The subjective scores ofimages in the NLB group were the highest (4.28 ± 0.74).Conclusion The NLB technique of dual-energy CT can improve the image quality of CTPA and reducethe radiation dose, providing more reliable imaging data for the clinical diagnosis of pulmonary embolism.

关 键 词:dual-energy computed tomography(CT) CT pulmonary angiography(CTPA) non-linear blending(NLB) image quality radiation dose 

分 类 号:R816.2[医药卫生—放射医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象