流动注射法在密云水库水中挥发酚测定中的适用性  

Adaptability of Flow Injection Method in the Determination of Volatile Phenol in Water of Miyun Reservoir

在线阅读下载全文

作  者:刘冬跃 王鑫 任修琳 魏一 刘金鑫 李心月 Liu Dongyue;Wang Xin;Ren Xiulin;Wei Yi;Liu Jinxin;Li Xinyue(Beijing Administration Office of Miyun Reservoir,Beijing 101512,China)

机构地区:[1]北京市密云水库管理处,北京101512

出  处:《山东化工》2023年第3期153-155,159,共4页Shandong Chemical Industry

摘  要:使用BDFIA-8000型号全自动流动注射分析仪对《水质挥发酚的测定流动注射-4-氨基安替比林分光光度法》进行方法验证。验证结果显示,标准曲线、检出限、测定下限、精密度、标准样品测定、加标回收率测定等指标均满足方法要求,结合考核断面比对结果,确保了该方法在密云水库水中挥发酚测定中的适用性。BDFIA-8000 automatic flow injection analyzer was used to verify the method of“Water Quality,Determination of Volatile Phenols,Flow Injection-4-Aminoantipyrine Spectrophotometer”.The verification results show that the standard curve,detection limit,determination lower limit,precision,standard sample determination,standard addition recovery determination and other indicators all meet the method requirements,combined with the comparison results of the examination sections,the adaptability of this method in the determination of volatile phenols in Miyun Reservoir water is ensured.

关 键 词:流动注射 挥发酚 方法验证 适用性 

分 类 号:O657.3[理学—分析化学]

 