血液透析机相关参数测量结果不确定度分析  被引量:2

Uncertainty Analysis of Measurement Results of Hemodialysis Machine Related Parameters

在线阅读下载全文

作  者:朱玉玲 ZHU Yuling(Tongren Inspection and Testing Institute,Tongren 554300,China)

机构地区:[1]铜仁市检验检测院,贵州铜仁554300

出  处:《品牌与标准化》2023年第3期37-39,共3页

摘  要:血液透析机是模拟人体肾功能,纠正体内电解质与维持酸碱度平衡的一组体外平衡装置,其相关参数测量结果的质量控制直接关系到患者的身体健康。本文通过GUM法对血液透析机电导率示值误差、温度示值误差、静脉压示值误差、流量示值误差的测量结果进行不确定度来源分析,并利用贝塞尔公式法、极差法和经验法等对相关参数的测量结果进行不确定度评定。Hemodialysis machines are a group of external balance devices that simulate human renal function,correct electrolytes in the body,and maintain acid-base balance.The quality control of the measurement results of their related parameters is directly related to the life and health of patients.This article uses the GUM method to analyze the sources of uncertainty in the measurement results of electrical conductivity indication error,temperature indication error,venous pressure indication error,and flow indication error of hemodialysis machines,and uses the Bessel formula method to analyze the sources of uncertainty.The range method and empirical method are used to evaluate the uncertainty.

关 键 词:血液透析机 测量结果 不确定度 

分 类 号:TH77[机械工程—仪器科学与技术]

 