检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:邵磊 王晗 刘昌杰 宋海蕴 张健 杨晨光 陈池来 SHAO Lei;WANG Han;LIU Chang-jie;SONG Hai-yun;ZHANG Jian;YANG Chen-guang;CHEN Chi-lai(Institute of Intelligent Machines,Hefei Institutes of Physical Science,Chinese Academy of Sciences,Hefei 230031,China;Science Island Branch,Graduate School of University of Science and Technology of China,Hefei 230026,China;Institute of Deep-sea Science and Engineering,Chinese Academy of Sciences,Sanya 572000,China)
机构地区:[1]中国科学院合肥物质科学研究院智能机械研究所,安徽合肥230031 [2]中国科学技术大学研究生院科学岛分院,安徽合肥230026 [3]中国科学院深海科学与工程研究所,海南三亚572000
出 处:《分析测试学报》2023年第6期749-754,共6页Journal of Instrumental Analysis
基 金:中国科学院战略先导专项(XDA22020503);中国科学院科研装备研制项目(YJKYYQ20210008);中国科学院合肥物质科学研究院院长基金(YZJJ202203-CX);国家重点研发计划(2021YFC2800301)。
摘 要:水下质谱是用于水中溶解气原位检测的重要手段,其定量标定依赖于制备不同浓度的溶解气,需要耗费大量时间。该文基于物理传输机制的不同,对溶解气过膜过程、管道传输过程以及分析器检测过程进行分解解析,提出了一种将3个过程分开标定的快速定量方法,并进行了实验验证。结果表明,过膜流量与样品分压高度线性,相关系数(r)为0.998,使用单组数据测定膜渗透效率与多组数据拟合结果的相对误差为3.9%;水下质谱大气阻器件管道会对过膜流量产生显著影响,不考虑膜后压强的传统模型存在极大误差,而考虑膜后压强的管道传输模型可将误差至少降低85.4%;质谱特征峰峰高与样品流量高度线性,相关系数均大于0.999。在以上结论的基础上,进一步使用鼓泡法制备不同浓度的溶解气对分步定量方法进行验证,得到分步标定结果与气相色谱仪检测结果的最大相对误差为10.9%。该文为时间受限情况下的水下质谱仪快速标定提供了一种新方法。Underwater mass spectrometry is an important means for in-situ detection of dissolved gas in water.Its quantitative calibration depends on the preparation of different concentrations of dis⁃solved gas,which takes a lot of time.Based on the different physical transport mechanisms,the process of dissolved gas passing through the membrane,pipeline transmission,and analyzer detec⁃tion was decomposed and analyzed in this paper,then a rapid quantitative method that separates the process of sample passing through the membrane from the mass spectrometry detection process was proposed,and finally an experimental verification was conducted.The results showed that the per⁃meation rate was highly linear with the partial pressure of the sample,with a correlation coefficient(r)of 0.998.Meanwhile,the membrane permeation efficiency could be determined by using a sin⁃gle set of data,and the relative error with the fitting results of multiple sets of data is 3.9%.Further⁃more,the pipeline has a significant impact on the permeation rate.The traditional model that does not consider the pressure behind the membrane has a large error,but the pipeline transmission model that considers the pressure behind the membrane could reduce the error by at least 85.4%.The peak heights of the characteristic peaks in the mass spectrum are highly linear with the sample flux,and their correlation coefficients are all greater than 0.999.Based on the above conclusions,the stepby-step quantitative method was verified by using the bubbling method to prepare different concentra⁃tions of dissolved gas,and the maximum relative error between the calibration results and the gas chromatograph detection results is 10.9%.Therefore,a new method for rapid calibration of underwa⁃ter mass spectrometers under time-limited conditions is provided.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.170