Oxylog3000plus便携式呼吸机在转运过程中耗氧量的测量与分析  被引量:1

Measurement and Analysis of Oxygen Consumption of Oxylog3000plus Portable Ventilator During Transportation

在线阅读下载全文

作  者:赖美红[1] 徐云飞 张子贤 LAI Meihong;XU Yunfei;ZHANG Zixian(Intensive Care Unit,Hangzhou Hospital of Traditional Chinese Medicine,Hangzhou Zhejiang 310012,China;Department of Medical Equipment,Hangzhou Hospital of Traditional Chinese Medicine,Hangzhou Zhejiang 310012,China)

机构地区:[1]杭州市中医院重症医学科,浙江杭州310012 [2]杭州市中医院医学装备部,浙江杭州310012

出  处:《中国医疗设备》2023年第7期34-38,共5页China Medical Devices

基  金:浙江省医药卫生科技计划项目(2019KY520)。

摘  要:目的 探究便携式呼吸机在院内转运中氧气配备的推算公式,为转运危重患者配备氧气源提供依据。方法 模拟转运过程,将Oxylog3000plus便携式呼吸机连接至模拟肺,在转运呼吸机氧气进气管串接气体质量流量计,记录转运中各个环节的氧气消耗值,并分析呼吸机运行过程中氧气消耗实测值与理想计算值、修正计算值和呼吸机监测值的相关性。结果 呼吸机运行阶段,氧气消耗的实测值、理想计算值、修正计算值和呼吸机监测值比较,差异均有统计学意义(P<0.001);测试呼吸机管路与模拟肺断开连接30 s,呼吸末正压为0、5、10 cmH_(2)O时断开连接过程中氧气消耗量比较,差异均有统计学意义(P<0.001)。Oxylog3000plus便携式呼吸机用于院内转运时氧气配备的推算公式为:转运呼吸机配备氧气源总量=氧气消耗的修正计算值/min×(估计转运时间+试用时间+预留时间)+断开连接30 s氧气消耗量+氧气钢瓶最低供气压力对应的残余氧气量。结论 便携式呼吸机在院内转运中氧气配备的推算公式可为危重患者院内转运配备合适氧气源、确保转运安全、减少氧气浪费提供一定的参考依据。Objective To explore the calculation formula for oxygen supply of portable ventilators during intra-hospital transportation,and provide a basis for oxygen supply for critical patients during transportation.Methods The transportation process was simulated by connecting the Oxylog3000plus portable ventilator to a simulated lung,and a gas mass flow meter was connected to the ventilator’s oxygen inlet pipe to record the oxygen consumption at each stage of transportation.The correlation between the measured oxygen consumption during the ventilator’s operation and the ideal calculated value,modified calculated value,and monitored value were analyzed.Results During the operation of the ventilator,the measured oxygen consumption,the difference between the measured value,ideal calculated value,corrected calculated value,and ventilator monitoring value of oxygen consumption was statistically significant(P<0.001).When the ventilator circuit was disconnected from the simulated lung for 30 s,there were statistically significant difference in oxygen consumption during disconnection when the positive end respiratory pressure was 0,5,and 10 cm H_(2)O(P<0.001).The calculation formula for oxygen supply of Oxylog3000plus portable ventilator during intrahospital transportation was:total oxygen supply of the ventilator=corrected calculated oxygen consumption value/min×(estimated transportation time+trial time+reserve time)+oxygen consumption during 30 s disconnection+residual oxygen volume corresponding to the minimum gas supply pressure of the oxygen cylinder.Conclusion The calculation formula for oxygen allocation during the transportation of portable ventilators in hospitals can provide a certain reference basis for equipping critically ill patients with appropriate oxygen sources,ensuring transportation safety,and reducing oxygen waste.

关 键 词:便携式呼吸机 院内转运 耗氧量 危重患者 

分 类 号:R197.39[医药卫生—卫生事业管理]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象