双能X射线骨密度仪的质量控制检测与质量管理  被引量:2

Quality Control Detection and Quality Management of Dual Energy X-Ray Absorptiometry

在线阅读下载全文

作  者:王猛[1] 王玲[1] 周翠红[2] 李环[1] WANG Meng;WANG Ling;ZHOU Cuihong;LI Huan(Department of Nuclear Medicine,China-Japan Friendship Hospital,Beijing 100029,China;Department of Anesthesiology,China-Japan Friendship Hospital,Beijing 100029,China)

机构地区:[1]中日友好医院核医学科,北京100029 [2]中日友好医院手术麻醉科,北京100029

出  处:《中国医疗设备》2023年第8期48-52,共5页China Medical Devices

基  金:中日友好医院科研基金(QN-30)。

摘  要:目的对双能X射线骨密度仪(Dual-Energy X-Ray Absorptiometry,DXA)进行质量控制检测,探讨DXA设备的质量控制检测与质量管理方法。方法使用美国GE-Lunar公司生产的DXA设备质量保证(Quality Assurance,QA)模块和脊椎模型,对DXA设备进行准确度误差和精确度误差的质量控制检测,并使用Minitab质量管理统计软件进行质量管理。结果DXA设备检测QA模块高、中、低骨密度(Bone Mineral Density,BMD)的准确度误差绝对值分别为0.10%、0.10%、0.40%,精确度误差分别为0.70%、0.70%、0.80%;DXA设备检测QA模块高、中、低BMD的休哈特(Shewhart)质量控制图上各检测点均在基线附近随机波动,均未偏离Shewhart法则;DXA设备检测脊椎模体L1~L4:BMD、骨矿物质含量、面积的累积和质量控制图上各标绘点均在中心线附近随机波动,且所有累积和线均在上下控制界限内。结论通过对DXA设备进行定期质量控制检测和质量管理,可实时掌握设备的工作状态,为临床及科研工作提供准确、可靠的检测结果。Objective To detect the quality control of dual-energy X-ray absorptiometry(DXA),and to discuss the quality control detection and quality management methods of DXA equipment.Methods Quality assurance(QA)module and spine phantom of DXA equipment produced by GE-Lunar Company were used to test the accuracy and precision errors of DXA equipment,and Minitab quality management statistical software was used for quality management.Results The absolute value of accuracy error of DXA equipment of high,medium and low bone mineral density(BMD)was 0.10%,0.10%and 0.40%respectively,and the precision error was 0.70%,0.70%and 0.80%respectively.The values of Shewhart quality control chart of high,medium and low BMD measured by DXA equipment fluctuated randomly near the baseline,and did not deviate from Shewhart rule.The cumulative sum quality control chart of BMD,bone mineral content and area were measured by DXA equipment on the water phantom of spine from L1 to L4.The plotted points on the quality control chart fluctuated randomly near the center line,and all cumulative sum lines were within the upper and lower control limits.Conclusion Through regular quality control testing and quality management of DXA equipment,the working status of the equipment can be grasped in real time,and accurate and reliable test results can be provided for clinical and scientific research work.

关 键 词:双能X射线骨密度仪 质量控制 准确度误差 精确度误差 质量管理 

分 类 号:R197.39[医药卫生—卫生事业管理]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象