摘 要:原子吸收分光光度计是一种用于微量或痕量分析的精密光学仪器,但是在使用过程中会出现不同问题和故障。基于此,介绍原子吸收分光光度计的结构和工作原理,分析造成原子分光光度计元素灯不能点亮的原因,并阐述了此种故障的解决办法。Atomic absorption spectrophotometer is a precise optical instrument used for trace or trace analysis.However,different problems and malfunctions may occur during use.This article mainly analyzes the solutions to the phenomenon of element lamps not being able to light up,and also introduces the structure and working principle of atomic absorption spectrophotometer.Its purpose is to conduct a more in-depth analysis of the reasons why the atomic spectrophotometer element lamp cannot be lit,and to elaborate on the solutions to this malfunction.