作 者:彭康强[1] 戴心仪 刘辉明[1] 何妮[1] 江琳玲 钟林昌 谢传淼[1] PENG Kangqiang;DAI Xinyi;LIU Huiming;HE Ni;JIANG Linling;ZHONG Linchang;XIE Chuanmiao(State Key Laboratory of Oncology in South China/Department of Medical Imaging,Sun Yat-sen University Cancer Center,Guangzhou 510060,China)
机构地区:[1]中山大学肿瘤防治中心影像科/华南恶性肿瘤防治全国重点实验室,广东广州510060
出 处:《中国医学物理学杂志》2023年第9期1091-1097,共7页Chinese Journal of Medical Physics
基 金:国家自然科学基金(82271300);广东省自然科学基金(2023A1515012739)。
摘 要:目的:比较自适应图像接收(AIR)线圈与传统乳腺专用线圈在乳腺磁共振(MR)成像中的应用,以期在保证图像质量的基础上提高受检者乳腺MR检查舒适度,为乳腺MR检查中线圈选择提供新方案。方法:对2023年1~3月在中山大学肿瘤防治中心行乳腺MR检查的34例健康女性受检者,分别采用两种射频线圈行乳腺MR检查。检查前后各发放调查问卷1份,收集受检者个人基本信息及检查舒适度情况;对比在两种线圈下T_(2)WI fs和MUSE-DWI序列中的图像信噪比、表观弥散系数值,并由2名高年资医师对图像质量进行主观评价;分析舒适度及图像质量,比较两种线圈在乳腺MR检查中的优劣。结果:AIR线圈舒适度优于乳腺专用线圈,差异有统计学意义(P<0.05);AIR线圈下T_(2)WI fs、MUSE-DWI序列信噪比均高于乳腺专用线圈图像,差异有统计学意义(P<0.05),两组图像表观弥散系数值之间,差异无统计学意义(P>0.05);AIR线圈下T_(2)WI fs序列主观图像质量低于乳腺专用线圈图像,差异有统计学意义(P<0.05),MUSE-DWI序列两组图像主观评价差异无统计学意义(P>0.05)。结论:乳腺MR检查中,AIR线圈舒适度优于传统乳腺专用线圈,具有一定的临床应用及推广价值。Objective To compare the applications of adaptive image receiving(AIR)coil and traditional coil in breast magnetic resonance imaging(MRI)for improving the comfort level of breast MRI on the basis of ensuring image quality,and providing a new selection of radiofrequency coils in breast MRI.Methods Atotal of 34 healthy female subjects who underwent breast MRI in Sun Yat-Sen University Cancer Center from January to March 2023 were enrolled in the study,and two kinds of radiofrequency coils were used for breast MRI.Through questionnaire survey before and after examination,the personal information was collected,and the comfort during breast MRI was investigated.AIR coil and traditional coil were used for MRI using T_(2)WI fat-suppressed sequence(T_(2)WI fs)and multiplexed sensitivity encoding diffusion-weighted imaging(MUSE-DWI)sequence.The signal-to-noise ratio(SNR)and apparent diffusion coefficient(ADC)value of the obtained images were compared,and the image quality was evaluated subjectively by two senior physicians.The advantages and disadvantages of two kinds of radiofrequency coils in breast MRI were discussed through the statistical analysis on comfort level and image quality.Results The comfort level of AIR coil was superior to that of traditional breast coil(P<0.05).The SNR of both T_(2)WI fs and MUSE-DWI sequences in AIR coil group was higher than that in traditional coil group(P<0.05),while there was no statistical difference inADC values between two groups(P>0.05).In terms of subjective evaluation,the image quality of T_(2)WI fs sequence in AIR coil group was lower than that in traditional coil group(P<0.05),while the difference in the subjective evaluation of MUSE-DWI sequence between two groups was trivial(P>0.05).Conclusion Compared with traditional coil,AIR coil for breast MRI has higher comfort level,worthy of popularization in clinic.
分 类 号:R318[医药卫生—生物医学工程] R816.4[医药卫生—基础医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.106