非接触式眼压计测量结果重复性的实验研究  被引量:1

Experimental Study on the Repeatability of Non-contact Tonometer Measurement Results

在线阅读下载全文

作  者:秦亭亭[1] 林正皓 张忠立[1] Qin Tingting;Lin Zhenghao;Zhang Zhongli(Shanghai Institute of Measurement and Testing Technology,Shanghai 201203)

机构地区:[1]上海市计量测试技术研究院,上海201203

出  处:《中国仪器仪表》2023年第8期36-39,共4页China Instrumentation

摘  要:本文首先介绍了非接触式眼压计的重复性检定方法,然后利用自主研发的基于标准模拟人眼的非接触式眼压计检定装置对六种不同品牌和型号的非接触式眼压计进行了重复性测量与数据分析。最后,针对检定过程中个别眼压计重复性超差的问题进行了测试与调整,并提出了相应的解决方法。This paper first introduces the repeatability verification method of non-contact tonometer,and then uses the independently developed non-contact tonometer verification device based on the standard simulated human eyes to carry out the repeatability measurement and data analysis of six different brands and models of non-contact tonometer.Finally,the problems of the repeatability of individual tonometers in the verification process are analyzed and adjusted,and the corresponding solutions are proposed.

关 键 词:非接触式眼压计 模拟人眼 重复性 测量 

分 类 号:TH786[机械工程—仪器科学与技术]

 