作 者:张景逍 张晗钰 张金彪[1] Zhang Jingxiao;Zhang Hanyu;Zhang Jinbiao(Cangzhou Hospital of Integrated TCM-WM,Heibei Province,Cangzhou Hebei 061000,China)
机构地区:[1]河北省沧州中西医结合医院,河北沧州061000
出 处:《医疗装备》2023年第18期90-94,共5页Medical Equipment
摘 要:目的根据XN1000-B4全自动血液分析仪中“PLT Clumps”的Q-Flag报警值判断低值血小板(PLT)结果的可靠性。方法通过对XN1000-B4全自动血液分析仪分析的低值PLT标本进行人工推片镜检,积累标本的Q-Flag报警值的分布及对应的PLT聚集情况,进行灵敏度和特异度分析并绘制ROC曲线。结果200份PLT聚集阳性标本的Q-Flag报警值分布于20~300之间。收集400份Q-Flag报警值>20的低值PLT标本进行人工推片镜检,结果显示,不同Q-Flag报警值分组的PLT聚集情况(阴性、阳性)比较,差异有统计学意义(χ^(2)=97.547,P=0.000);且随着Q-Flag报警值的增高,PLT聚集的阳性率增高,趋势性χ^(2)检验显示差异有统计学意义(趋势性χ^(2)=0.494,P=0.000);当Q-Flag报警值≥250时,PLT聚集阳性的灵敏度为95.0%。收集Q-Flag报警值<30的200份低值PLT标本进行人工推片镜检,结果显示,Q-Flag报警值<20时标本未发生PLT聚集的特异度为100.0%。ROC曲线提示,以Q-Flag报警值预测PLT聚集情况的最佳临界值为100,此时曲线下面积为0.900,灵敏度为86.5%,特异度为75.6%。结论通过XN1000-B4全自动血液分析仪“PLT Clumps”的Q-Flag报警值判断低值PLT标本是否发生PLT聚集具有较高的可行性,其应用不仅利于检验工作者提高工作效率,而且更好地保证了检验质量。Objective The reliability of low value platelet(PLT)results was determined based on the Q-Flag alarm value of"PLT Clumps"in the XN1000-B4 automatic blood analyzer.Methods Through manual slide microscopic examination of low value PLT samples analyzed by XN1000-B4 automatic hematology analyzer,the distribution of Q-Flag alarm values and corresponding PLT aggregation of samples were accumulated.In addition,sensitivity and specificity analysis was analyzed,and the ROC curve was drawn.Results The Q-Flag alarm values of 200 PLT aggregation positive specimens ranged from 20 to 300.With collection of 400 low value PLT samples with Q-Flag alarm values greater than 20 for manual slide microscopy,the results showed that there was a statistically significant difference in PLT aggregation(negative and positive)between different Q-Flag alarm value groups(χ^(2)=97.547,P=0.000);In addition,the positive rate of PLT aggregation increased with the increase of Q-Flag alarm value,and the trendχ^(2)test showed a statistically significant difference(trendχ^(2)=0.494,P=0.000);When the Q-Flag alarm value was≥250,the sensitivity of PLT aggregation positive was 95.0%.With collection of 200 low-value PLT specimens with Q-Flag alarm value<30 for artificial microscope examination,the results showed that the specificity of specimens without PLT aggregation with Q-Flag alarm values<20 was 100.0%.The ROC curve indicated that the optimal critical value for predicting PLT aggregation with the Q-Flag alarm value was 100.At this time,the area under the curve was 0.900,the sensitivity was 86.5%,and the specificity was 75.6%.Conclusions The Q-Flag alarm value of XN1000-B4 automatic blood analyzer"PLT Clumps"has a high feasibility in determining whether PLT aggregation occur in low-value PLT specimens.Its application is not only conducive to improving the work efficiency of laboratory workers,but also better ensures the quality of examination.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.222