作 者:王国鹏 WANG Guopeng
机构地区:[1]广东省惠州市质量计量监督检测所
出 处:《计量与测试技术》2023年第9期118-120,共3页Metrology & Measurement Technique
摘 要:医用电动止血带属于手术室、急诊室、诊疗室设备,主要用于手术室供阻断人体肢体血流用。其中压力传感器是其计量溯源的主要内容。近年来,对医用电动止血带的校准需求越来越多,但目前并未有相关的研究。本文通过研究医用电动止血带的校准方法,并对其自动泄压功能、校准方法和校准进行不确定度评定,为医疗计量工作提供参考。Medical electric tourniquet belongs to the operating room,emergency room,consultation room equip⁃ment,mainly used in the operating room for blocking the blood flow of human limbs.Pressure sensor is the main content of its metrological traceability.In recent years,there is an increasing demand for the calibration of medical electric tourniquet,but there is no relevant research at present.This paper studies the calibration method of medical electric tourniquet,and evaluates the uncertainty of its automatic pressure relief function,calibration method and calibration,so as to provide reference for medical metrology.
关 键 词:医用电动止血带 时间检查 自动泄压检查 压力示值误差的校准 不确定度分析