检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:吴思圻 洪宝玉[1] 胡志雄[1] 马雪然 卞昕[1] 刘文丽[1] WU Si-qi;HONG Bao-yu;HU Zhi-xiong;MA Xue-ran;BIAN Xin;LIU Wen-li(National Institute of Metrology,Beijing 100029,China)
出 处:《计量学报》2023年第9期1473-1478,共6页Acta Metrologica Sinica
基 金:国家重点研发计划(2016YFF0201001)。
摘 要:基于气腹机工作原理及相关计量技术规范开展气腹机校准方法的研究,建立模拟腹腔系统,搭建气腹机计量检测装置。重点论述气腹机计量检测装置的软硬件功能架构,采用自动化测量程序实现了气腹机的智能化计量校准。针对一款典型的气腹机,开展计量校准检测并进行不确定度评定,气压设置值误差扩展不确定度为77.02 Pa(k=2),流量显示值误差扩展不确定度为5.98%(k=2),结果表明研制的气腹机计量检测装置能够应用于气腹机性能检测。Based on the working principle of insufflator and relevant measurement technical specifications,the research on the calibration method of insufflator is carried out,a simulated abdominal cavity system is established,and the calibration device of insufflator is built.Software and hardware functional architecture of the calibration device of insufflator are elaborated,intelligent calibration of insufflator is realized by utilizing automatic measurement program.A typical insufflator model is calibrated and the uncertainty is evaluated.The expanded uncertainty of air pressure set value error is 77.02 Pa(k=2),and the expanded uncertainty of flow display value error is 5.98%(k=2).The results show that the developed calibration device of insufflator can be applied to the performance testing of insufflator.
关 键 词:计量学 气腹机 计量检测装置 不确定度 智能化校准
分 类 号:TB99[一般工业技术—计量学] TB935[机械工程—测试计量技术及仪器]