新华牌PS-100X型过氧化氢低温等离子灭菌器故障维修8例  

Maintenance of Shinva PS-100X Medical Low-Temperature Hydrogen Peroxide Plasma Sterilizer:8 case reports

在线阅读下载全文

作  者:达尼亚尔·麦海提 李明际 郭海平 阿丽耶 Daniyaer Maihaiti;LI Ming-ji;GUO Hai-ping;Aliye(Department of Medical Engineering of Medical Support Center,the 947th Hospital of the PLA Army,Shule 844200,Xinjiang Wei Autonomous Region,China)

机构地区:[1]陆军第947医院医疗保障中心医学工程科,新疆疏勒844200

出  处:《医疗卫生装备》2023年第9期117-120,共4页Chinese Medical Equipment Journal

摘  要:阐述了新华牌PS-100X型过氧化氢低温等离子灭菌器的结构和工作原理,介绍了该设备出现的8例故障现象,分析了故障产生的原因,并给出了具体的排除方法,指出了该类设备日常保养和定期检查的重要性,为临床工程人员维修同类故障提供了参考。The structure and working principle of Shinva PS-100X Medical Low-Temperature Hydrogen Peroxide Plasma Sterilizer were introduced.Eight cases of the sterilizer failures had their phenomena,causes and elimination described in detail.The importance of daily maintenance and periodical examination was pointed out for the sterilizer.References were provided for clinical engineers to treat similar failures.

关 键 词:新华牌PS-100X型低温等离子灭菌器 故障分析 故障排除 

分 类 号:R318.6[医药卫生—生物医学工程] TH771.4[医药卫生—基础医学]

 