基于FSE序列的MRI设备性能测试的探讨  

Discussion on Performance Test of MRI Equipment Based on FSE Sequence

在线阅读下载全文

作  者:高林[1] 方旭超 刘红光 李常虹 张行坤 GAO Lin;FANG Xuchao;LIU Hongguang;LI Changhong;ZHANG Xingkun

机构地区:[1]青岛大学附属青岛市中心医院

出  处:《计量与测试技术》2023年第11期66-68,71,共4页Metrology & Measurement Technique

摘  要:本文对国家新制定的《医用磁共振系统校准规范(征求意见稿)》进行验证;分析并探讨快速自旋回波序列(FSE)代替自旋回波序列(SE),对磁共振设备性能测试的可行性;依据征求意见稿的测试条件与计算方法,对该设备进行性能测试,并分别采用自旋回波序列(SE)和快速自旋回波序列(FSE)扫描SMR-170模体,对成像参数进行对比分析。实验证明:在征求意见稿指定的性能测试参数条件下,SE序列对设备的性能测试具有可行性。In this paper,the newly formulated"Medical Magnetic Resonance System Calibration Standard(draft for comment)"is verified;The feasibility of using fast spin echo sequence(FSE)instead of spin echo sequence(SE)to test the performance of magnetic resonance equipment is analyzed and discussed.According to the test conditions and calculation methods of the draft,the performance of the device was tested,and the spin echo sequence(SE)and fast spin echo sequence(FSE)were used to scan the SMR-170 module,and the imaging parameters were compared and analyzed.Experiments show that the SE sequence is feasible for the performance test of the equipment under the conditions of the performance test parameters specified in the draft.

关 键 词:磁共振成像 模体 性能测试 自旋回波序列 快速自旋回波序列 

分 类 号:R445.2[医药卫生—影像医学与核医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象