检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:卢俊平 何春泽 霍贞 王洁 LU Junping;HE Chunze;HUO Zhen;WANG Jie(Facilities and Support Center,Academy of Military Medical Sciences,Beijing 100850,China)
机构地区:[1]军事医学研究院科研保障中心,北京100850
出 处:《化学分析计量》2023年第12期95-98,共4页Chemical Analysis And Meterage
摘 要:介绍实验室pH(酸度)计的校准和正确使用与维护保养的方法。首先依据能特斯方程讨论了实验室pH计的单点校准、两点校准和三点校准的校准原理和适用范围,然后讨论了邻苯二甲酸氢钾、磷酸氢二钠和磷酸二氢钾、硼砂这三种pH计校准用标准物质在三种校准方法下pH值随温度变化的情况,结果表明,两点校准较单点校准可获得更准确可靠的测试结果。最后介绍了实验室pH计在日常使用中电极的活化、清洗等日常维护保养方法。The calibration,correct use and maintenance method of laboratory pH meters were introduced.Firstly,the calibration principles and applicable ranges of single-point calibration,two-point calibration and three-point calibration for laboratory pH(acidity)meters were discussed in detail according to Nentz equation,and then,the pH changes of three standard substances of KHC8H4O4,Na2HPO4-KH2PO4,and Na2B4O7·10H2O for pH meter calibration were explored with the temperature by using three calibration methods in experiment,the test results showed that two point calibration were more accurate and reliable than single-point calibration.Finally,the maintenance methods of the laboratory pH(acidity)meter,such as electrode activation and cleaning,were introduced in daily use.
关 键 词:实验室pH(酸度)计 酸度计 校准 维护保养