基于多波长近红外光谱的血红蛋白浓度无创检测技术研究  

Research on non-invasive detection technology of hemoglobin concentration based on multi-wavelength near-infrared spectroscopy

在线阅读下载全文

作  者:彭福来 陈财 张宁玲 王星维 吕丹阳 王卫东[2] PENG Fulai;CHEN Cai;ZHANG Ningling;WANG Xingwei;L Danyang;WANG Weidong(Shandong Zhongke Advanced Technology Co.,LTD.,Jinan 250000,China;Medical Device R&D and Clinical Evaluation Center,The General Hospital of the People's Liberation Army,Beijing 100853,China)

机构地区:[1]山东中科先进技术有限公司,济南250000 [2]中国人民解放军总医院医疗器械研发与临床评价中心,北京100853

出  处:《生物医学工程研究》2024年第1期70-75,共6页Journal Of Biomedical Engineering Research

基  金:山东省自然科学基金资助项目(ZR2020QF024,ZR2021ZD40)。

摘  要:针对传统的血红蛋白浓度检测需要抽血采样,检测流程复杂且无法连续监测血红蛋白浓度的变化趋势等问题,本研究设计了一种基于多波长近红外光谱的无创血红蛋白浓度检测方法。首先,基于Beer-Lambert定律建立了血红蛋白无创检测数学模型,并依据该模型设计了八波长近红外光电容积脉搏波(photoplethysmography,PPG)信号采集系统;然后,对采集的PPG信号进行降噪和滤除基线漂移等预处理,并根据建立的无创检测模型对特征信息进行提取与选择;最后,基于Stacking算法构建血红蛋白预测回归模型。通过对249例临床数据进行实验验证,得到无创检测模型的预测值与参考值的均方根误差为1.17 g/dL,相关系数为0.75。实验结果表明,本研究方法可有效实现血红蛋白浓度的无创检测。Aim at the problem that traditional hemoglobin concentration detection requires blood sampling,the detection process is complicated and cannot continuously monitor the variation trend of human hemoglobin concentration,we proposed a non-invasive hemoglobin detection method based on multi-wavelength near-infrared spectroscopy.Firstly,a mathematical model for non-invasive hemoglobin detection was established based on the Beer-Lambert law,and an eight-wavelength near-infrared PPG signal acquisition system was designed based on this model.Then,the collected PPG signals were pre-processed to reduce noise and baseline drift,and the feature information was extracted and selected according to the established non-invasive detection model.Finally,the hemoglobin prediction regression model was constructed based on the Stacking algorithm.249 clinical data were used to verify the performance,the root-mean-square error between the predicted value of the non-invasive detection model and the reference value was 1.17 g/dL,and the correlation coefficient was 0.75.The experimental results show that this method can achieve non-invasive detection of hemoglobin concentration effectively.

关 键 词:血红蛋白浓度 无创检测 PPG信号处理 Stacking回归模型 

分 类 号:R318[医药卫生—生物医学工程]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象