粉末压片-X射线荧光光谱法测定富硅土壤和沉积物样品中的5种重金属元素  被引量:2

Determination of 5 Heavy Metal Elements in Silicon-rich Soils and Sediments by X-ray Fluorescence Spectrometry with Pressed Powder Pellet

在线阅读下载全文

作  者:陈春霏 卢秋 姚苏芝 梁晓曦 洪欣 李丽和 韦江慧 CHEN Chunfei;LU Qiu;YAO Suzhi;LIANG Xiaoxi;HONG Xin;LI Lihe;WEI Jianghui(Guangxi Zhuang Autonomous Region Ecological and Environmental Monitoring Centre,Nanning,Guangxi 530028,China)

机构地区:[1]广西壮族自治区生态环境监测中心,南宁530028

出  处:《中国无机分析化学》2024年第5期513-520,共8页Chinese Journal of Inorganic Analytical Chemistry

基  金:广西自然科学基金资助项目(2021GXNSFAA075028)。

摘  要:X射线荧光光谱法测定土壤和沉积物中的重金属具有简便、快速、准确等优点,但现有的环境行业标准方法不适用于SiO 2含量大于80%的土壤和沉积物的测定。为提高X射线荧光光谱法测定土壤和沉积物中的重金属在环境监测领域中的适用性,建立基于粉末压片-X射线荧光光谱法测定富硅土壤和沉积物中重金属的分析方法。通过选取富硅的土壤和沉积物标准物质提高校准曲线的测量范围以及优化各元素的测量条件,探讨基体效应和谱线重叠干扰、研究压片制样的最佳压力等途径,测定富硅土壤和沉积物中的Pb、Cr、Cu、Ni、Zn。结果表明,在已优化的测定条件下,通过延长校准曲线的测量上限,保证待测元素含量在校准曲线范围内,同时采用经验系数法和康普顿散射内标法校正基体效应,可进一步提高方法的准确度和精密度。方法比对结果显示,X射线荧光光谱法的测定结果与原子吸收光谱法的测定结果无显著性差异。选用2种不同的富硅标准物质进一步验证方法的精密度和准确度,Pb、Cr、Cu、Ni、Zn的测定结果与标准物质认定值的对数误差为0.002~0.08,相对标准偏差为1.0%~4.6%。方法的精密度和正确度满足土壤和水系沉积物环境监测分析的技术要求,具有良好的应用前景。X-ray fluorescence spectrometry has the advantages of simple,rapid and accurate for determination of heavy metals in soils and sediments.However,the existing environmental standard method is not suitable for the determination of soils and sediments with SiO 2 content more than 80%.In order to improve the applicability of heavy metals determination in soils and sediments by X-ray fluorescence spectrometry in environmental monitoring,a method based on X-ray fluorescence spectrometry with pressed powder pellet was established for the determination of heavy metals in silicon-rich soil and sediment samples.Pb,Cr,Cu,Ni and Zn in silicon-rich soils and sediments were determined by selecting silicon-rich standard materials to improve the measurement range of calibration curve and optimize measurement conditions of each element,discussing the correction of matrix and spectral line overlap interference,and studying the optimal pressure of sample preparation.The results showed that under the optimized measurement conditions,the accuracy and precision of the method could be further improved by extending the measurement upper limit of the calibration curve to ensure that the content of the elements were within the range of the calibration curve.Meanwhile,the empirical coefficient method and compton scattering internal standard method were adopted to correct the matrix effect for further improving the accuracy and precision of the method.The method comparison results showed that there was no significant difference between the measurement results of X-ray fluorescence spectroscopy and atomic absorption spectrometry.Two different silicon-rich standard materials were used to verify the precision and accuracy of the method.The logarithmic error between the measured results and the certified results of standard materials was 0.002-0.08 and the relative standard deviation was 1.0%-4.6%.The precision and accuracy of the method meet the analysis requirements of soil and stream sediment in environmental monitoring area,and has a good a

关 键 词:X射线荧光光谱法 富硅土壤和沉积物 重金属元素 粉末压片法 

分 类 号:O657.34[理学—分析化学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象