电感耦合等离子体质谱(ICP-MS)法测定生物制药用一次性储液瓶中34种可提取无机元素  

Determination of 34 Extractable Inorganic Elements in Single-use Bottles Used in Biopharmaceutical Manufacturing by ICP-MS

在线阅读下载全文

作  者:高杰 陈星 张玲帆[2] 陈佳 朱海伟 邢春秀 GAO Jie;CHEN Xing;ZHANG Lingfan;CHEN Jia;ZHU Haiwei;XING Chunxiu(Bio-Link Inspection&Testing(Jiangsu)Co.,Ltd.,Nantong,Jiangsu 226500,China;Research Center of Analysis and Test,East China University of Science and Technology,Shanghai 200237,China)

机构地区:[1]百林科生物检测(江苏)有限公司,江苏南通226500 [2]华东理工大学分析测试中心,上海200237

出  处:《中国无机分析化学》2024年第5期607-614,共8页Chinese Journal of Inorganic Analytical Chemistry

基  金:国家自然科学青年基金资助项目(21407050)。

摘  要:为建立同时测定生物制药用一次性储液瓶中34种可提取无机元素含量的方法,采用5种溶剂对样品分别提取1、21和70 d,利用电感耦合等离子体质谱仪(ICP-MS)检测其中的元素含量,通过比较各检出元素的含量与人每日最大允许暴露量(PDE)进行风险评估。结果表明,各元素拟合曲线线性良好,相关系数≥0.999,不同提取溶剂中,方法的检出限为0.001~25.8μg/L,相对标准偏差范围为0.20%~2.3%,低、中、高加标平均回收率在86.0%~116%。实际样品测定时,不同提取条件下可提取元素的含量在水和碱性溶液中较高,而随提取时间的增加趋势并不显著。各检出元素的含量均远低于其PDE值。方法检出限低,线性良好,具有较高的准确度和精密度,可以满足生物制药用一次性储液瓶中多种可提取无机元素同时检测的需求。To establish a method for simultaneous determination of 34 extractable elements content in single-use bottles used in biopharmaceutical manufacturing,five solvents were used to extract the samples for 1,21 and 70 days,respectively.Element contents was detected using inductively coupled plasma mass spectrometry(ICP-MS),and risk assessment was performed by comparing each detected element content with the maximum permitted daily exposure(PDE)of individuals.The results showed that the fitting curves of each element were well,with the corresponding liner correlation coefficients of each element greater than 0.999.The limit of detection in different extraction solvents was 0.001-25.8μg/L,and the relative standard deviation(RSD)was 0.20%-2.3%.The average recovery of low,medium and high spiked samples were ranged from 86.0%-116%.For the actual sample determination,the element content of extractables under different extraction conditions in water and alkaline solution was higher,and the trend was not significant with extraction time increasing.The content of each element was much lower than its PDE value.The method has low detection limit,good linearity,and high accuracy and precision,which could meet the needs of simultaneous detection of multiple elements in single-use bottles for biopharmaceuticals.

关 键 词:一次性储液瓶 电感耦合等离子体质谱法 无机元素 可提取物 

分 类 号:O657.63[理学—分析化学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象