国产新型高压氧专用电控气动呼吸机运行稳定性的研究  被引量:1

Study on the operation stability of a new type domestic electro-pneumatic ventilator in hyperbaric oxygen chamber

在线阅读下载全文

作  者:王丛 薛连璧 于秋红[1] 刘亚玲[1] 任梓齐 刘影[1] Wang Cong;Xue Lianbi;Yu Qiuhong;Liu Yaling;Ren Ziqi;Liu Ying(Department of Hyperbaric Oxygen,Beijing Tiantan Hospital,Beijing 100070,China;Department of Hyperbaric Oxygen,Huazhong University of Science and Technology Union Shenzhen Hospital,Shenzhen,Guangdong 518052,China)

机构地区:[1]首都医科大学附属北京天坛医院高压氧科,北京100070 [2]华中科技大学协和深圳医院高压氧科,深圳1518052

出  处:《中华航海医学与高气压医学杂志》2024年第1期96-100,共5页Chinese Journal of Nautical Medicine and Hyperbaric Medicine

摘  要:目的探讨国产新型高压氧(HBO)专用电控气动呼吸机在高气压环境中以压力控制模式(PCV)运行时的稳定性。方法将呼吸机气动部分与模拟肺连接安装于舱内,呼吸机电控部分与模拟肺数据采集电脑置于舱外,在PCV模式下根据预设吸气压力(PI)分5个预设水平(1.0、1.5、2.0、2.5、3.0 kPa),按舱内压力分为0.10 MPa组、0.15 MPa组、0.20 MPa组,采集气道峰压(P_(peak))、潮气量(V_(T))和分钟通气量(MV);每组20个循环,比较各预设PI在3组中的参数的变化。结果随环境压力升高,各组预设PI中P_(peak)、V_(T)、MV均出现下降,差异均有统计学意义(均P<0.05),下降幅度<6%。结论国产高压氧专用电控气动呼吸机在HBO舱内(0.10~0.20 MPa)以PCV运行时随着舱内压力升高呼吸参数出现差异,但该变化在临床使用相对稳定范围内。Objective To investigate the operation stability of a new type domestic electro-pneumatic ventilator during pressure-controlled ventilation(PCV)in hyperbaric oxygen(HBO)chamber.Methods The pneumatic part of the HBO ventilator and a test lung were connected and placed inside a hyperbaric chamber.The electric part of the ventilator was placed outside the chamber along with a computer connected with the test lung for real-time data collection.During the PCV,the inspiratory pressures(PI)were preset at five levels(1.0,1.5,2.0,2.5,and 3.0 kPa),under three different chamber pressures(0.10 MPa,0.15 MPa,and 0.20 MPa),respectively.The peak inspiratory pressure(P_(peak)),the tide volume(V_(T)),and the minute volume(MV)as displayed by the ventilator and the test lung were recorded at each level of PI for 20 cycles.The differences of the three groups of data were then compared and analyzed.Results As the chamber pressure increased,the P_(peak),V_(T),and the MV at each level of PI decreased within a narrow range(<6%),and the differences were statistically significant(P<0.05).Conclusion When the HBO electro-pneumatic ventilator working in PCV mode in a HBO chamber(0.10–0.20 MPa),its ventilation indexes can have some changes as the pressure increasing,but the changes are still within the relative stable range for clinical application.

关 键 词:高压氧 呼吸机 间歇正压通气 呼吸力学 气道阻力 

分 类 号:R197.39[医药卫生—卫生事业管理]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象