作 者:舒宇康 赵潮 杜小玉 何湘 赵航 史晓凤[1] 马君[1] Shu Yukang;Zhao Chao;Du Xiaoyu;He Xiang;Zhao Hang;Shi Xiaofeng;Ma Jun(Institute of Physics and Optoelectronic Engineering,Ocean University of China,Qingdao 266100,China)
机构地区:[1]中国海洋大学物理与光电工程学院,山东青岛266100
出 处:《红外与激光工程》2024年第4期266-275,共10页Infrared and Laser Engineering
基 金:山东省重点研发计划项目(2019GHY112027)。
摘 要:对谱域光学相干层析系统(SD-OCT)进行成像深度扩展,能清晰地探测到更深层次的样品信息。提出结合去除复共轭镜像和焦平面叠加的方法来提高探测深度。首先通过光束偏移振镜中心,对样品添加载波频率,基于希尔伯特变换滤除负频信号,去除复共轭镜像,再对样品不同深度聚焦成像,实现聚焦平面的叠加,最终扩展成像深度。基于去复共轭成像系统测量不同样品的图像,验证了该方法的有效性和可靠性。针对透明片材料,系统有效成像范围扩大一倍,从1.97 mm达到3.94 mm,针对结肠组织,有效成像深度从0.51 mm扩展到0.65 mm。结果表明,深度扩展技术能够提升成像范围,同时也突破了焦深限制,样品深层结构信息也清晰可见。Objective Spectral domain optical coherence tomography(SD-OCT)is a technique used for tomography and three-dimensional imaging of the internal microstructure of materials and biological tissues.By performing Fourier transform on the collected interference signal,the SD-OCT system can reconstruct image depth information,providing a detailed view of the sample's structure.However,the resulting image includes a mirror image with zero phase delay symmetry,which can be problematic.To address this issue,the sample is positioned on the zero phase delay side during scanning to eliminate the influence of the mirror image.This approach,while effective,only utilizes half of the imaging range space.Improving the detection depth of OCT systems by eliminating complex conjugate images is crucial for enhancing the accuracy and reliability of the imaging process.Methods The deconjugated imaging system was developed from the original SD-OCT system(Fig.3).The system utilizes a broadband light source,with the light being split by a beam splitting cube into the reference arm and sample arm.The reference arm consists of a collimator and a reflector,while the sample arm includes a twodimensional galvanometer,scanning lens,and sample stage.The two-dimensional galvanometer enables scanning of the sample in both X and Y directions by deflecting the reflector.The entire sample arm system is finely tuned using an electronic displacement platform to adjust the relative position,create beam offset,and add carrier frequency.The sample stage can be moved axially to focus imaging at different depths within the sample,allowing for the superposition of focusing planes and enhancing the imaging depth.Results and Discussions To eliminate the influence of the complex conjugate image,the distance between the beam and the center of the mirror is adjusted to 1.2 mm.Before the spectral data undergoes the standard image reconstruction process,the complex conjugate image is removed using the added carrier frequency through Hilbert transform(Fig.5).This res
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.12.149.233