作 者:刘子宁 刘鑫[1] LIU Zi-ning;LIU Xin(Medical Apparatus and Equipment Deployment,Zhujiang Hospital of Southern Medical University,Guangdong Guangzhou 510280)
机构地区:[1]南方医科大学珠江医院设备器材科,广东广州510280
出 处:《中国医疗器械信息》2024年第10期158-160,164,共4页China Medical Device Information
摘 要:VIVID E9是具有全新设计的D系列探头的彩色多普勒超声波诊断仪,由美国GE公司2008年推出,全新设计的D系列探头,能够实现快速的成像图像传输。但该设备长时间使用经常会出现无法开机、黑屏或者死机等问题,本文分析了VIVIDE9彩色超声诊断仪容易出现的故障,探讨其根治维修难题的方法。VIVID E9 is a color Doppler ultrasonic diagnostic instrument with a newly designed D-series probe,which was launched by GE in 2008 in the United States.The newly designed D-series probe can achieve fast image transmission,which is conducive to the inspection of vascular disease and reproductive organ diseases.However,long-term use of this device often leads to problems such as inability to start,black screen,or crash.This article analyzes the faults that are prone to occur with the VIVID E9 color ultrasound diagnostic instrument and explores the methods to solve the maintenance problems.
关 键 词:彩色超声诊断仪 VIVIDE9 图像干扰 故障维修 启动异常 散热
分 类 号:R197.39[医药卫生—卫生事业管理]