泰安台水管仪标定故障典型实例  

Typical Example Analysis of Calibration Fault of Water Pipe Instrument in Tai'an Station

在线阅读下载全文

作  者:李惠玲[1] 林秀娜[1] 张明[1] 于庆民[1] 赵小贺[2] 赵帝 LI Huiing;LIN Xiuna;ZHANG Ming;YU Qingmin;ZHAO Xiaohe;ZHAO Di(Tai'an Seismic Monitoring Center Station of Shandong Earthquake Agency,Tai'an 271000,China;Linyi Seismic Monitoring Center Station of Shandong Earthquake Agency,Linyi 276000,China)

机构地区:[1]山东省地震局泰安地震监测中心站,山东泰安271000 [2]山东省地震局临沂地震监测中心站,山东临沂276000

出  处:《华北地震科学》2024年第2期89-93,共5页North China Earthquake Sciences

基  金:中国地震局震情跟踪定向工作任务(2024010208)。

摘  要:针对泰安台DSQ水管仪在2020年的年中、年终标定中出现NS向多次重复标定均不合格的情况,通过分析标定过程中单端电压差值的变化,判断为标定装置中的水银掉落导致N端堵塞,经检修后标定合格。此次典型故障排查中分析电压差变化规律的方法,对台站水管仪的运行维护、标定故障识别有很好的借鉴和指导意义。The calibration of Tai'an DSQ water pipe instrument in the year-mid and year-end of 2020,there were many repeated calibrations in NS direction,which were unqualified.Analysis of the change of voltage difference in single terminal during calibration process,it is judged that the mercury in the calibration device fell,resulting in N-terminal blockage.The calibration is qualified after maintenance.In the process of typical fault investigation,the method of analyzing the variation law of single-ended voltage difference has a good reference and guiding significance for the operation and maintenance of station water pipe instrument and the identification of calibration fault.

关 键 词:水管仪 标定故障 电压差 泰安台 