Imaging plate scanners calibration and the attenuation behavior of imaging plate signals  

在线阅读下载全文

作  者:薄楠 王乃彦 Nan Bo;Nai-Yan Wang(College of Nuclear Science and Technology,Beijing Normal University,Beijing 100875,China;China Institute of Atomic Energy,Beijing 102413,China)

机构地区:[1]College of Nuclear Science and Technology,Beijing Normal University,Beijing 100875,China [2]China Institute of Atomic Energy,Beijing 102413,China

出  处:《Chinese Physics B》2024年第6期38-44,共7页中国物理B(英文版)

基  金:supported by the Nuclear Industry Academician Fund, the National Natural Science Foundation of China (Grant No. 1220051312);Young Talents Fund of China National Nuclear Corporation (Grant No. FY212406000901)。

摘  要:Based on previously reported work, we propose a new method for calibrating image plate(IP) scanners, offering greater flexibility and convenience, which can be extended to the calibration tasks of various scanner models. This method was applied to calibrate the sensitivity of a GE Typhoon FLA 7000 scanner. Additionally, we performed a calibration of the spontaneous signal attenuation behavior for BAS-MS, BAS-SR, and BAS-TR type IPs under the 20±1℃ environmental conditions, and observed significant signal carrier diffusion behavior in BAS-MS IP. The calibration results lay a foundation for further research on the interaction between ultra-short, ultra-intense lasers and matter.

关 键 词:image plate SCANNER ultra-short ultra-intense lasers 

分 类 号:TN249[电子电信—物理电子学]

 