三级公立医院除颤监护仪质量检测实践与改进措施  被引量:1

Quality Control Testing Practice and Improvement Measures of Defibrillation Monitor in Prefectures and Cities Public Tertiary Hospitals

在线阅读下载全文

作  者:唐局 郑瑞强[1,2] 张勤 仲辉[1,2] 马宪礼 成定胜[1,2] 陆正大 薛谭[1,2] TANG Ju;ZHENG Ruiqiang;ZHANG Qin;ZHONG Hui;MA Xianli;CHENG Dingsheng;LU Zhengda;XUE Tan(Northern Jiangsu People’s Hospital;Yangzhou University Hospital Management Institute,Yangzhou,Jiangsu,225001)

机构地区:[1]江苏省苏北人民医院 [2]扬州大学医院管理研究院,江苏扬州225001

出  处:《江苏卫生事业管理》2024年第7期980-984,共5页Jiangsu Health System Management

基  金:江苏省医院协会医院管理创新研究课题(JSYGY-3-2023-632);苏北人民医院管理课题(YY⁃GL202214)。

摘  要:目的:探讨除颤监护仪的质量控制检测方法,提高其使用安全性和有效性。方法:采用美国福禄克(FLUKE)专业检测设备对某地市级公立医院45台在用除颤监护仪的外观、开机、电气安全、能量精度、充电时间、同步延迟时间和心率误差共7个项目进行检测。结果:45台除颤监护仪初检合格率73%。其中外观合格率93%,开机合格率75%,能量精度合格率97%,电气安全、充电时间、同步延迟时间和心率误差合格率均为100%。针对不合格问题提出改进措施,纠正后二次检测全部合格。结论:对除颤监护仪进行质量控制检测可有效降低其使用安全风险,对区域内医疗设备质量控制管理工作具有指导意义。Objective:The objective of this study was to discuss the quality control testing method for defibrillation monitors and enhance their safety and effectiveness.Methods:FLUKE professional testing equipment was used to assess the appearance,power-up functionality,electrical safety,energy accuracy,charging time,synchronization delay time,and heart rate error of 45 active defibrillation monitors in a certain municipal public hospital.Results:The overall pass rate for the 45 defibrillation monitors was found to be 73%.Specifically,the appearance qualified rate was 93%,boot qualified rate was 75%,energy accuracy qualified rate reached 97%,while electrical safety,charging time,synchronization delay time and heart rate error all achieved a perfect score of 100%.Based on identified causes for non-compliance issues during initial testing phase,improvement measures were implemented resulting in successful retesting outcomes.Conclusion:Quality control testing for defibrillation monitors can effectively mitigate usage-related safety risks and provide valuable guidance for medical equipment quality control and management within the region.

关 键 词:公立医院 除颤监护仪 质量控制 改进措施 

分 类 号:R197.39[医药卫生—卫生事业管理]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象