X射线荧光光谱法测定超基性岩中14种主次量元素  

Determination of 14 Elements in Ultrabasrocks by X-ray Fluorescence Spectroscopy

在线阅读下载全文

作  者:艾力尼亚孜·吐尔逊 张淼 Alniyaz·Tursun;Zhang Miao(The Mineral Experimental Research Institute of Xinjiang Uygur Autonomous Region,Xinjiang Key Laboratory of Rock and Mineral Analysis and Process Mineralogy,Urumqi 830000,China)

机构地区:[1]新疆维吾尔自治区矿产实验研究所,新疆岩石矿物分析及工艺矿物学研究重点实验室,新疆乌鲁木齐830000

出  处:《广东化工》2024年第14期149-150,175,共3页Guangdong Chemical Industry

摘  要:本论文确定了X射线荧光熔片法测定超基性岩中Si、Ca、Fe、Ni、Co、V等14种主次量元素的方法。样品与助溶剂比例为1∶10,样品在700℃预氧化,1050℃预熔2 min,旋转、摇摆10 min,铂-黄金坩埚自然冷却制备玻璃样片。用国家一级标准物质和人工混合标样制备校准标准样片,建立校准曲线,用熔片法大大降低了矿物效应和粒度效应,用经验系数法补偿基体效应,检测元素除Co相关系数r=0.99665外,其他元素相关系数均大于0.999,本方精密度、准确度法满足化学分析要求。The method of measuring Si,Ca,Fe,Ni,Co,and V in X-ray fluorescence melting method was determined.The ratio of sample to cosolvent was 1∶10,the samples were pre-oxidized at 700℃,premelted at 1050℃for 2 min,rotated and rocking for 10 min,and naturally cooled with platinum-glass crucible.The calibration standard sample is prepared by national level standard material and artificial mixed standard sample,and the calibration curve is established,the mineral effect and particle size effect is greatly reduced by melting method,the matrix effect is compensated by empirical coefficient method,except Co correlation coefficient r=0.99665,the correlation coefficient is greater than 0.999.The precision and accuracy method meet the requirements of chemical analysis.

关 键 词:超基性岩 X射线荧光 熔片法 主次量元素 

分 类 号:TQ[化学工程]

 