等离子体发射光谱仪代替火焰光度计测定玉米全钾含量的可行性分析  

Feasibility analysis of plasma emission spectrometer instead of flame photometer to determine the content of total potassium in maize

在线阅读下载全文

作  者:高飞 范珊珊 王艳平 王伊琨 季卫 GAO Fei;FAN Shanshan;WANG Yanping;WANG Yikun;JI Wei(Beijing Cultivated Land Construction Protection Center,Beijing,100101,China)

机构地区:[1]北京市耕地建设保护中心,北京100101

出  处:《质量安全与检验检测》2024年第4期7-11,共5页QUALITY SAFETY INSPECTION AND TESTING

摘  要:本文用硫酸和过氧化氢消煮玉米的籽粒、棒轴、秸秆后,尝试用等离子体发射光谱仪代替火焰光度计进行样品全钾含量的测定,并对2种方法测定结果进行分析。结果表明,2种方法测定结果成显著线性相关。且等离子体发射光谱仪测定玉米中全钾含量结果精确度和加标回收率结果均较好,同时等离子体发射光谱仪具有检测速度快、操作简便的特点,为玉米样品全钾含量的测定提供参考。After the corn kernels,shaft and straw were cooked with sulfuric acid and hydrogen peroxide,the content of total potassium in the samples was determined by icp-aes instead of flame photometer,the results of the two methods were analyzed.The results showed that there was a significant linear correlation between the two methods.The results of determination of total potassium content in corn by icp-aes showed that the precision and recovery rate were both good,it can provide reference for the determination of total potassium content in corn samples.

关 键 词:等离子体发射光谱仪 火焰光度计 玉米 全钾 

分 类 号:S1[农业科学—农业基础科学]

 