基于荧光免疫法的地表水中2,4-D原位快速检测  

In-situ Rapid Detection of 2,4-D in Surface Water by Fluorescence Immunoassay

在线阅读下载全文

作  者:谭凌智 张琦 文幼成 朱圣清 翟婉盈 TAN Lingzhi;ZHANG Qi;WEN Youcheng;ZHU Shengqing;ZHAI Wanying(Changjiang Basin Ecology and Environment Monitoring and Scientific Research Center,Changjiang Basin Ecology and Environment Administration,Ministry of Ecology and Environment,Wuhan,Hubei 430010,China)

机构地区:[1]生态环境部长江流域生态环境监督管理局生态环境监测与科学研究中心,湖北武汉430010

出  处:《环境监测管理与技术》2024年第5期56-59,共4页The Administration and Technique of Environmental Monitoring

基  金:国家重点研发计划基金资助项目(2017YFF0108500);长江生态环境保护修复联合研究基金资助项目(2022-LHYJ-02-0507-05)。

摘  要:采用荧光免疫法原位测定地表水中2,4-二氯苯氧乙酸(2,4-D),方法在7.2μg/L~211.7μg/L范围内线性良好,最低检出限为2.6μg/L。对空白水、长江水和湖泊水样品做2,4-D的4个质量浓度水平的加标回收试验,7次测定结果的RSD为3.9%~5.8%,加标回收率为85.9%~106%。用该方法与国标法同时测定黄石市内某长江水源地和某湖泊水中的2,4-D,结果无明显差异。In this study,2,4-Dichlorphenoxyacetic acid(2,4-D)in surface water was determined in suit by fluorescence immunoassay.The method had good linearity in the range of 7.2μg/L to 211.7μg/L.The detection limit was 2.6μg/L.The spiked recoveries on blank and practical samples from Yangtze River and a lake at four mass concentration levels of 2,4-D were from 85.9%to 106%,and the RSDs of seven measurements were from 3.9%to 5.8%.There was no significant difference between the results by this method and the national standard method in determining 2,4-D in water from Yangtze River water source and a lake in Huangshi.

关 键 词:2 4-二氯苯氧乙酸 荧光免疫法 原位监测 快速检测 地表水 