检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:胥海亮 陈诚 王亮[1] 朱珠 郭旭影 王茜[1] 粟琨璞 XU Hailiang;CHEN Cheng;WANG Liang;ZHU Zhu;GUO Xuying;WANG Qian;SU Kunpu(Sichuan Radiation Environment Management and Monitoring Center Station,Chengdu 611100;Key Laboratory of Mass Spectrometry Analysis Technology of China National Nuclear Corporation,Chengdu 611130)
机构地区:[1]四川省辐射环境管理监测中心站,成都611100 [2]中核集团质谱分析技术重点实验室,成都611130
出 处:《辐射防护》2024年第5期461-468,共8页Radiation Protection
基 金:四川省生态环境保护科技项目应用基础研究类课题资助(2022HB03)。
摘 要:超低本底流气式正比计数器是样品总α/β测量的理想探测器,但本身具有较强的串道现象,影响样品测量的准确性。系统研究了流气式正比计数器在同一质量厚度、不同241 Am和40 K质量分数下α→β、β→α串道影响,建立了总α/β串道修正的二元方程,并与遮盖修正法进行对比,对自制的25组α/β模拟粉末样品进行了测量和串道修正。结果显示α→β串道影响较大,修正因子随着α计数率的变化并非恒定值;β→α串道修正因子影响较小,其平均值为0.001;遮盖修正法对模拟粉末样品中β测量结果的修正效果较差,其相对偏差绝对值介于1.4%~45.0%;二元方程修正法对模拟粉末样品中β测量结果的修正效果较好,其相对偏差绝对值介于0~5.5%。The ultra-low background flow gas proportional counter is the ideal detector for totalα/βsample measurement,but itself has strong crosstalk phenomenon,which affects the accuracy of sample measurement.This paper systematically studies the flow type proportional counter under the same mass thickness,different 241 Am and 40 K mass fractionsα→β,β→αCross channel influence.This paper established a binary equation for totalα/βchannel correction and compared it with the cover correction method and with the self-made 25 groupsα/βsimulate powder samples for measurement and channel correction.The results show that:The impact ofα→βcrosstalk is significant;The change withαcounting rate is not constant;The influence of theβ→αchannel correction factor is relatively small,with an average value of 0.001.The accuracy of the measurement results is poor by using the cover correction method to correct the simulated powderβsample,with absolute relative deviation ranging from 1.4%to 45.0%.The binary equation correction method is used to correct the simulated powderβsample.The accuracy of the measurement results is relatively high,with an absolute value of relative deviation ranging from 0 to 5.5%.
关 键 词:超低本底流气式正比计数器 串道修正 二元方程修正法
分 类 号:TL811.2[核科学技术—核技术及应用]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.137.161.247