检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:张猛 刘洋 盛训超 汤玉美 胡源[1] Zhang Meng;Liu Yang;Sheng Xunchao;Tang Yumei;Hu Yuan(State Key Laboratory of Fire Science,University of Science and Technology of China,Hefei 230026,Anhui,China;Hefei Zhongke Environmental Monitoring Technology National Engineering Laboratory Co.,Ltd.,Hefei 230088,Anhui,China;State Key Laboratory of Pulsed Power Laser Technology,National University of Defense Technology,Hefei230037,Anhui,China)
机构地区:[1]中国科学技术大学火灾科学国家重点实验室,安徽合肥230026 [2]合肥中科环境监测技术国家工程实验室有限公司,安徽合肥230088 [3]国防科技大学脉冲功率激光技术国家重点实验室,安徽合肥230037
出 处:《光学学报》2024年第18期323-330,共8页Acta Optica Sinica
基 金:国家重点研发计划(2022YFC3003100);安徽省重点研究与开发计划(2022a05020019);中国科学技术大学重要方向培育基金(WK2320000054)。
摘 要:阐述了基于气体滤波相关的非分散红外(NDIR)检测技术。首先,依据NDIR技术原理建立仿真模型,利用模型仿真构建系统核心件参数与探测信号之间的关系,得出参考光路和测量光路的光强比值,为后续放大电路设计提供参考。其次,通过系统响应函数评估系统的测量误差最大不超过1×10^(-6),并定量评估外界温度变化对系统造成的误差。外界温度变化10℃会导致反演的CO_(2)气体体积分数最大变化约为9×10^(-6),研究表明,为了保证仪器的精度和漂移要求,对长光程测量气室进行控温是必要的。所研制的NDIR分析仪使用模拟参数,并采用恒温技术将测量气室的温度控制在(45±0.1)℃,NDIR分析仪的测试结果表明,仪器检出限达到0.075×10^(-6),示值误差为0.19%,精密度(相对标准偏差)为0.11%,24 h内零点漂移低于0.033%ξ_(FS)(ξ_(FS)为满量程),且量程漂移也低于0.3%ξ_(FS)。NDIR技术测试结果和激光技术测试结果显著相关,其线性拟合度R^(2)为0.94。实际设计的仪器测试结果表明,该仿真方法具有一定的应用价值。Objective Reliable gas detection is essential for industrial control,health,and environmental protection.Gas detection based on the infrared absorption principle offers high selectivity but faces challenges in precision and stability.Non-dispersive infrared(NDIR)and gas filter correlation(GFC)analyzer are pivotal for precise gas detection among various measurement devices.In recent years,infrared technology has rapidly advanced due to efforts from major research institutions,companies,and universities.This study establishes a model to describe the relationship between the infrared light source,wavelength,GFC wheel,center wavelength,filter bandwidth,the optical path length of the air chamber,gas volume fraction,and the measurement/reference signals.This model provides insights for amplifier circuit design.We analyze the measurement accuracy and the influence of temperature variations on the system using the response function of the analyzer.Our design proposal enhances primary design stages,guiding the development of NDIR and GFC analyzer and demonstrating the practical application of our approach.Methods The NDIR and GFC analyzer comprises six main components:infrared light source,GFC wheel,filter,air chamber,detector,and main circuit system(Fig.1).To optimize and evaluate our design proposal,we develop a model to describe their relationships with measurement and reference signals.Firstly,we model the infrared light source,GFC wheel,center wavelength,filter bandwidth,the optical path length of the air chamber,and gas volume fraction,deriving expressions for measurement and reference signals.MATLAB simulations based on the HITRAN spectra database are employed to simulate NDIR absorption under varying gas volume fraction,temperature,pressure,and other conditions(Fig.3),providing insights for amplifier circuit design.We further optimize our design proposal by analyzing the system’s measurement accuracy through the response function(Fig.4).Simulations also assess gas absorption under different temperatures,quantify
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:18.218.241.211