Construction and investigation of all-inone microneedles complexed with functionalized polydiacetylene liposomes for improved in situ detection sensitivity  

在线阅读下载全文

作  者:Jiarui Wang Weimiao Li Yan Chen Meng Wu Yan Shi Jumin Lee Ming Kong 

机构地区:[1]College of Marine Life Sciences,Ocean University of China,Shandong,China [2]Department of Materials Science and Engineering,Pohang University of Science and Technology,Gyeongsangbuk-do,Republic of Korea [3]Qingdao Technical College,Shandong,China

出  处:《hLife》2024年第10期514-526,共13页健康科学(英文)

基  金:funded by the Natural Science Foundation of Shandong Province(ZR2022MC033);the Qingdao Science&Technology Benefiting People Project(24-1-8-smjk-11-nsh).

摘  要:Polydiacetylene(PDA)liposomes have been widely applied for detection due to their distinctive optical properties.How-ever,the liquid phase in which PDA liposomes are dispersed generates several drawbacks,for instance,instability,compromise of detection sensitivity induced by dilution,and separation of target sampling and detection,making it inconvenient for application.In this paper,various functionalized PDA liposomes for detecting target were prepared,which were also immobilized into swelling microneedles to construct a solid-phase detection system.The PDA lipo-somes-complexed microneedles(PDA/MNs)enable the integration of target sampling and detection in one platform.The effects of the dispersing matrix phase on the detection sensitivity of PDA liposomes were systematically investi-gated from both environmental and chemical perspectives.PDA/MNs exhibited higher sensitivity than their counter-parts in liquid phase.PDA/MNs were optimized and validated for lead ion(Pb^(2+))and sialic acid(SA)detections.For Pb^(2+) detection,the limit of detection(LOD)of the PDA/MNs was 13.7 mM and 2.5 times lower than the liquid phase.For SA detection,the LOD of the PDA/MNs was 0.83 mM and 1.7 times lower than the liquid phase.The results suggested that such PDA/MNs were validated to provide a label-free,stable,sensitive,and convenient tool in an all-in-one manner for physiologic target detection.

关 键 词:polydiacetylene liposomes complexed swellable microneedles ALL-IN-ONE improved detection sensitivity 

分 类 号:TH77[机械工程—仪器科学与技术]

 