作 者:邹海蓉 ZOU Hairong(Siemens Shanghai Medical Equipment Ltd.,Shanghai,201318)
机构地区:[1]上海西门子医疗器械有限公司,上海市201318
出 处:《中国医疗器械杂志》2024年第6期613-617,共5页Chinese Journal of Medical Instrumentation
摘 要:在DR系统设计中,高压发生器和球管是两个核心组件,它们的性能直接影响着系统的曝光效能。该文首先阐述了为何在DR产品设计过程中需要对这两个部件的性能进行深入评估,接着详细介绍了评估方法,包括单次曝光能力和连续曝光能力的评估。最后,该文提供了一些建议和措施,以提升连续曝光能力。这些方法和建议不仅有助于选择合适的关键部件,也为DR系统的设计、集成和测试提供了重要的参考依据。In the design of DR systems,high-voltage generators and X-ray tubes are two key components whose performance directly affects the system's exposure capabilities.This article first explains why it is necessary to conduct an in-depth evaluation of the performance of these two parts during the design of DR products.Then,it provides a detailed description of the evaluation methods,including the assessment of single exposure capability and continuous exposure capability.Finally,the article offers some suggestions and measures to enhance continuous exposure capability.These methods and suggestions not only help in selecting appropriate key components but also provide an important reference for the design,integration,and testing of DR systems.
分 类 号:R197.39[医药卫生—卫生事业管理] TH774[医药卫生—公共卫生与预防医学]