检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:李翔 雷强 史昌兵 郝爽雨 李晓婷 LI Xiang;LEI Qiang;SHI Changbing;HAO Shuangyu;LI Xiaoting(Hangzhou TongJian Technology Co.,Ltd.,Hangzhou 310000,Zhejiang,China;Zhejiang Institute of Medical Device Testing,Hangzhou 310000,Zhejiang,China)
机构地区:[1]杭州通鉴科技有限公司,浙江杭州310000 [2]浙江省医疗器械检验研究院,浙江杭州310000
出 处:《工程与试验》2024年第4期94-97,共4页Engineering and Test
摘 要:针对电子内窥镜设备电磁兼容测试中一次性通过率较低的辐射发射项目,对某电子内窥镜产品进行缺陷分析,采取相应的整改措施使产品满足试验测试要求,为电子内窥镜设备电磁兼容整改提供了参考。Aiming at the radiation emission test with low one-time pass rate in electromagnetic compatibility test of electronic endoscope equipment,the defect analysis of an electronic endoscope product is carried out.The corresponding rectification measures are taken to ensure that the product can meet the test requirements,which provides a reference for the electromagnetic compatibility rectification of electronic endoscope equipment.
分 类 号:TH776.1[机械工程—仪器科学与技术]