Principle verification of the calibration light source subsystem for the calorimeter in herd experiment  

在线阅读下载全文

作  者:Qianjun Chen Xin Liu Zhigang Wang Zheyuan Zhang Li Zhang Yunpeng Lu Guozheng Nie Kejun Zhu Yongwei Dong 

机构地区:[1]Hunan University of Science and Technology,Xiangtan,411201,China [2]Institute of High Energy Physics,Chinese Academy of Sciences(CAS),Beijing,100049,China

出  处:《Radiation Detection Technology and Methods》2024年第4期1664-1671,共8页辐射探测技术与方法(英文)

基  金:supported by the National Natural Science Foundation of China(Grant No.12027803).

摘  要:Purpose The High energy cosmic-radiation detection facility(HERD)is a space astronomy and particle astrophysics experiment planned to be installed on the Chinese space station.The core detector of HERD,a three-dimensional imaging calorimeter,uses LYSO crystals as the detection medium,with wavelength-shifting fibers readout by image intensifier and CMOS imaging chips.In order to monitor the performance of the calorimeter in orbit,a highly integrated calibration light source subsystem is required to test the pulse linearity of the readout system,including the optical fibers.Method To meet this critical requirement,a Calibration Light Source Subsystem was proposed based on the dual-light-source method,and a test circuit was set up to prove the principle.Results and conclusions Using a photomultiplier tube and waveform sampling module,the entire system’s linear response range was measured,which is equivalent to 1–900 Minimum Ionizing Particles(MIPs),with a linearity better than±5%.After adding an attenuator(transmittance of 6%)to the front end of the photomultiplier tube,the linear response range expands to 14–28,000 MIPs with a linearity better than±5%.This result demonstrated the significant contribution of the PMT’s saturation to the linearity measurement.The addition of the attenuator reduces the adverse effects of PMT saturation,and therefore the linearity of the Calibration Light Source Subsystem itself is better than±5%within the required dynamic range.In addition,testing of a single light source showed that long-term stability can be maintained within a range of±0.5%under conditions of 10–20℃.The calibration method described in this paper has the characteristics of high integration and independent control of channels,providing an efficient and flexible calibration solution for the readout of calorimeter in HERD experiment.

关 键 词:HERD CALORIMETER Calibration light source subsystem Dual-light-source method 

分 类 号:O57[理学—粒子物理与原子核物理]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象