检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:方奕文[1] 贾丽[1] 余林[1] 宋一兵[1] 孙长勇[1]
机构地区:[1]汕头大学化学系分析测试中心,广东汕头515063
出 处:《光谱学与光谱分析》2003年第1期73-74,共2页Spectroscopy and Spectral Analysis
基 金:汕头大学工业催化学科"2 11工程"建设项目资助
摘 要:采用消化法预处理血清样品 ,并用ICP AES法测定样品中的Se ,同时 ,探索Fe的存在对测定的干扰情况以及干扰的消除方法。结果表明 ,Se的检测限为 0 0 0 76μg·mL- 1 ,加标回收率为 91 6 % ,1 0次测定的RSD(n =1 0 )为 2 1 4 7% ;在样品中加入KOH ,可以大大降低Fe对Se测定的干扰。The serum sample solution was prepared by using acid digestion and the Se concentration in it was determined by ICP-AES. At the same time, the interference from Fe and the interference-free method were studied. The results showed that the recovery of standard addition was 91.6%, the relative standard deviation(n=10)was 2.147% and when KOH was added to the sample solution the interference was reduced greatly. The method is efficient, accurate and easy to operate.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.36