呼吸机加呼气末正压时误触发的原因和对策  被引量:1

The causes of error trigger when mechanical ventilation using PEEP and its countermeasure.

在线阅读下载全文

作  者:蔡若夫[1] 陈爱君[1] 

机构地区:[1]浙江大学医学院附属第一医院,310003

出  处:《现代医药卫生》2003年第4期391-392,共2页Journal of Modern Medicine & Health

摘  要:目的:对临床上几款呼吸机加呼气末正压(PEEP)时的误触发分析其原因和相应的对策。方法:主要对在不同的呼吸机上20例故障根据其使用年限进行故障归类。结果:发生呼气阀和周边管路漏气故障65%,传感器(压力和流量)的损坏25%,电磁阀的老化10%,生物医学工程的故障排除是处理的实用方法。结论:呼气调和周边管路漏气,传感器损坏及电磁阀老化是3种主要故障。Objective: To analyze the causes of error trigger about several types of respiratory machine when their mechanical ventilations were used with PEEP and to bring forward the countermeasure.Methods: To classify 20 faults of error trigger in different machines according to the usage years.Results: The principal faults of error trigger are: leakage of expiratory valve and circuit(65%), damage of sensor(25%) and aging of electromagnetic valve(10%) .Conclusion:To bring forward the practical measures to deal with the three principal faults.

关 键 词:呼吸机 呼气末正压 误触发 原因 对策 

分 类 号:R459.6[医药卫生—治疗学]

 